Velleman DVM8852 User Manual Download Page 6

DVM8852 

00 (04/11/2008) 

 

Velleman

®

 

6

 

The meter has two operational modes, fast or slow. In fast mode, sound is sampled every 
125ms; in slow mode every second. Select fast or slow mode with the mode select button 

[E]

Selected mode is indicated in the display, 

[5]

 or 

[6]

. Note that the display itself is updated 2 

times per second. 

 

There are two weighing schemes available on the meter: A-weighted (dBA) and C-weighted 
(dBC): 

ƒ

 

dBA

: sound pressure adjusted towards the frequency range of human hearing 

ƒ

 

dBC

: sound pressure adjusted towards the low frequency end of the spectrum 

Select the desired scheme with the signal weighing select button 

[F]

, the scheme is indicated on 

the display 

[18]

 or 

[19]

 

Press the hold button 

[J]

 to freeze the current measurement on the display. Hold mode is 

indicated on the display 

[21]

. Press the hold button 

[J]

 again to exit hold mode. 

 

Press the max/min capture button 

[G]

 to continuously display the maximum measured value, 

display indicates 

MAX [3]

. Press the max/min capture button 

[G]

 again to continuously display 

the minimum measured value, display indicates 

MIN [4]

. Pressing the max/min capture button 

[G] 

again will show the current noise pressure level again. 

 

Press the backlight button 

[K]

 to switch the backlight on and off. 

 

The auto-power off mode indication 

[14]

 is shown when auto-power off is enabled. The meter 

switches off after ±15 minutes of user inactivity. To disable the auto-power off function, press 
the setup button 

[D]

 

To switch off the meter, press and hold the power on/off button 

[I]

 for about 3 seconds (display 

shows countdown). 

 

The meter can be used handheld or mounted on the included tripod. Screw the tripod into the 
mounting hole 

[P]

 on the back of the meter. 

Note:

 the AC/DC output jack can be used to connect other devices e.g. frequency analyzer (not 

incl.), plotter (not incl.) … 

 

AC output 

voltage: 1Vrms/range step 
impedance: 100

 

DC output  voltage: 10mV/dB 

impedance: 1k

 

Ground  

Time/date setting

 

 

If the meter is on, switch it off first. 

 

Press and hold the setup button 

[D]

 while switching on the meter with the power on/off button 

[I]

. Release the setup button 

[D] 

when 

TIME [12]

 is displayed. The display now shows the 

date.  

 

Pressing the setup button 

[D] 

to enter the setup menu. Press the setup button 

[D]

 to save the 

setting and go to the next setup entry. Press the hold button 

[J]

 at any time to save the setting 

and exit setup mode. 

Press # 

Display 

 

minutes 

TIME 

               n n

 

set the minutes by pressing the range/level select button 

[H]

 

hour 

TIME 

h-A

 or 

h-P

 

set the hour by pressing the range/level select button 

[H]

 

A

=ante meridiem (before noon), 

P

=post meridiem (after noon) 

3 DATE 

–d- 

day 

set the day of the month by pressing the range/level select button 

[H]

 

DATE –H- 

month

 

set the month by pressing the range/level select button 

[H]

 

5 DATE 

–Y- 

year

 

set the year by pressing the range/level select button 

[H]

 

6 rSt 

press the hold button 

[J]

 to reset date and time to factory default 

Note:

 the day is not shown on the display during normal operation. It can only be shown by going 

into setup mode 

Data logger function 

 

During measurement, press the record start/stop button 

[C]

 to start recording the 

measurement. Press again to stop recording. Recordings can be transferred to a computer (see 

§7 Analysis). 

 

To clear the memory, switch off the meter first. Press and hold the record start/stop button 

[C]

 

while switching the meter back on. Release the record start/stop button 

[C]

 when the word 

CLR

 

appears on the display. 

Warning:

 all recorded data in the meter is lost. Consider uploading the 

memory content to a computer prior to clearing the memory.

 

Summary of Contents for DVM8852

Page 1: ...TA LOGGER GELUIDSDRUKMETER DATALOGGER SONOM TRE ENREGISTREUR DE DONN ES SCHALLPEGELMESSGER T DATENLOGGER SON METRO DATA LOGGER USER MANUAL 4 GEBRUIKERSHANDLEIDING 9 NOTICE D EMPLOI 14 MANUAL DEL USUAR...

Page 2: ...DVM8852 00 04 11 2008 Velleman 2 Figure 1 Figure 2 Figure 3...

Page 3: ...DVM8852 00 04 11 2008 Velleman 3 Figure 4 Figure 5...

Page 4: ...y from rain moisture splashing and dripping liquids Protected the meter against extreme heat and dust Do not use in area s where altitude 2 000m Damage caused by disregard of certain guidelines in thi...

Page 5: ...mit 9 bar graph scale depends on selected range low medium high auto 10 DATE only displayed during setup 11 18 88 88 shows time during normal operation shows set up codes during set up 12 TIME indicat...

Page 6: ...f the meter Note the AC DC output jack can be used to connect other devices e g frequency analyzer not incl plotter not incl X AC output voltage 1Vrms range step impedance 100 Y DC output voltage 10mV...

Page 7: ...o a histogram Go to the Graph G menu select Line and click anywhere in the graph to change the appearance back to line Also in zoom mode double click left mouse button on a point of the graph will set...

Page 8: ...life 30 hours with alkaline battery dimensions 278 x 76 x 50mm weight 350g auto power off 15min operation temperature 0 40 C operation humidity 90 included accessories instruction manual tripod batter...

Page 9: ...r tegen stof en extreme temperaturen Deze geluidsdrukmeter is niet geschikt voor gebruik 2 000 m De garantie geldt niet voor schade door het negeren van bepaalde richtlijnen in deze handleiding en uw...

Page 10: ...aag gemiddeld hoog automatisch 10 DATE datum enkel tijdens instelling 11 18 88 88 tijdsaanduiding tijdens normaal gebruik instelcodes tijdens instelling 12 TIME tijdsaanduiding 13 188 8 gemeten waarde...

Page 11: ...r P Opmerking Sluit een extern toestel zoals een frequentieanalyzer niet meegelev plotter niet meegelev aan de meter via de the AC DC uitgang N X AC uitgang spanning 1 V rms stap impedantie 100 Y DC u...

Page 12: ...en histogram over te schakelen Keer terug naar de bargraph door het menu Graph G te openen de optie Line te selecteren en in de histogram te klikken In zoommodus klik dubbelklik met de linkermuisknop...

Page 13: ...ensduur batterij 30 uur met alkalinebatterij afmetingen 278 x 76 x 50 mm gewicht 350 g automatische uitschakeling 15 min werktemperatuur 0 40 C vochtigheidsgraad 90 meegeleverde accessoires handleidin...

Page 14: ...temp ratures extr mes Ce sonom tre ne convient pas une utilisation une altitude de plus de 2 000 m La garantie ne s applique pas aux dommages survenus en n gligeant certaines directives de cette notic...

Page 15: ...gamme 9 graphique chelle en fonction de la gamme s lectionn e bas moyen haut automatique 10 DATE ne s affiche que lors du param trage 11 18 88 88 affichage de l heure en mode normal affichage des code...

Page 16: ...graphique non incl la sortie CA CC X sortie CA tension 1 V RMS cran imp dance 100 Y sortie CC tension 10 mV dB imp dance 1 k Z masse Param trage de l heure de la date teindre le sonom tre Maintenir en...

Page 17: ...er en un histogramme Pour r afficher le graphique d rouler le menu Graph G s lectionner l option Line et cliquer dans l histogramme En mode zoom cliquer double dans le graphique pour tablir une ligne...

Page 18: ...0 heures avec pile alcaline dimensions 278 x 76 x 50 mm poids 350 g d sactivation automatique 15 min temp rature de service 0 40 C taux d humidit 90 accessoires fournis notice d emploi tr pied pile to...

Page 19: ...extremas No es apto para un uso a una altura de m s de 2 000m Los da os causados por descuido de las instrucciones de seguridad de este manual invalidar n su garant a y su distribuidor no ser responsa...

Page 20: ...edici n lenta cada 1 s 7 88 indicaci n del rango l mite inferior 8 188 indicaci n del rango l mite superior 9 gr fico de barras el gr fico de barras depende del rango seleccionado bajo medio elevado a...

Page 21: ...cuenta atr s Es posible instalar el son metro en un tr pode Atornille el tr pode en la conexi n P de la parte trasera del aparato Nota Conecte un analizador de frecuencias no incl un bode plotter no...

Page 22: ...eleccione la opci n Histogram Haga clic en el gr fico de barras para cambiarlo en un histograma Para volver a visualizar el gr fico de barras abra el men Graph G seleccione la opci n Line y haga clic...

Page 23: ...nom a 30 horas con pila alcalina dimensiones 278 x 76 x 50 mm peso 350 g desactivaci n autom tica 15 min temperatura de funcionamiento 0 40 C humedad 90 accesorios incluidos manual del usuario tr pode...

Page 24: ...zwasser aus Sch tzen Sie das Ger t vor Staub und extremen Temperaturen Dieses Schallpegelmessger t eignet sich nicht f r Anwendung 2 000 m Bei Sch den die durch Nichtbeachtung der Bedienungsanleitung...

Page 25: ...ngt vom ausgew hlten Bereich niedrig durchschnittlich hoch automatisch ab 10 DATE Datum nur w hrend Einstellung 11 18 88 88 Uhranzeige w hrend einer normalen Anwendung Einstellcodes w hrend Einstellu...

Page 26: ...haltet Schalten Sie die Funktion mit der Einstellungstaste D ein und aus Halten Sie den EIN AUS Schalter I 3 Sekunden gedr ckt um das Schallpegelmessger t auszuschalten Display z hlt ab Sie k nnen das...

Page 27: ...en Bargraph ber das Men Real Time R Mehr Details werden ber dem Bargraph angezeigt Stoppen Sie das Echtzeit Monitoring wenn Sie die Abtastfrequenz einstellen m chten Diese Abtastfrequenz ist NICHT die...

Page 28: ...ed 50 100 dB Hi 80 130 dB Genauigkeit 1 4 dB ref 94 dB 1 kHz Frequenzbewertung A C Messbereiche schnell 125ms langsam 1s dynamischer Bereich 50 dB Frequenzbereich 31 5 Hz 8 kHz Mikrofon Elektret Konde...

Reviews: