5 Connecting to power supply
VEGAWAVE 62 • Two-wire 8/16 mA
32255-EN-181126
1
2
4
3
5
5
5
5
Fig. 9: Material versions, single chamber housing
1 Plastic (not with Ex d)
2 Aluminium
3 Stainless steel (not with Ex d)
4 Stainless steel, electropolished (not with Ex d)
5 Filter element for pressure compensation or blind plug with version IP 66/
IP 68, 1 bar (not with Ex d)
For connection to a signal conditioning instrument. The sensor is
powered via the connected signal conditioning instrument. For further
information see the "
Technical data
" in the "
Supplement
".
The wiring example is applicable for all suitable signal conditioning
instruments.
If the mode switch of VEGAWAVE 62 is correctly set to "max.", the
control lamp on VEGAWAVE 62 lights.
•
red - with submerged vibrating element
•
green - with uncovered vibrating element
Take note of the operating instructions manual of the signal condition-
ing instrument. Suitable signal conditioning instruments are listed in
chapter "
Technical data
".
1
Fig. 10: Wiring plan
1 Voltage supply
Housing overview
Wiring plan