15
4 Mounting
VEGAMIP R62 • Receiving unit
41911-EN-200522
Type of material
Window material
Max. thickness
Glass or ceramic
Inspection glass
< 200 mm
Aluminiumoxide ceram-
ic Al
2
O
3
< 200 mm
Abrasion protection (ce-
ramic)
< 100 mm
Isolating material
Mineral wool
< 1000 mm
Ceramic fibre plates
< 1000 mm
Fireclay bricks
< 500 mm
A
C
B
Fig. 8: Installation in front of a vessel window
A Window of non-conductive material, for example, glass, plastic, etc.
B Mounting with condensation on the inner wall of the vessel
C Inspection glass fitting
In general, the same as for metal vessels applies to concrete vessels
due to the steel braining.
In general, the same as for metal vessels applies to wood vessels due
to the liquid absorption of the wood.
VEGAMIP 62 can be used for detection of products in pipelines. In
pipelines of non-metallic products such as plastic or glass, the meas-
ured product can be detected through the pipe wall.
Concrete vessel
Wood vessels
Pipelines