56
13 Supplement
VEGABAR 38 • Three-wire with IO-Link (2 x transistor or 4 … 20 mA plus 1 x transistor)
57532-EN-191014
Operating voltage U
B
- illuminated display
and adjustment unit
15 … 35 V DC
Reverse voltage protection
Integrated
Permissible residual ripple
Ʋ
for U
N
12 V DC (12 V < U
B
< 18 V)
≤ 0.7 V
eff
(16 … 400 Hz)
Ʋ
for U
N
24 V DC (18 V < U
B
< 35 V)
≤ 1 V
eff
(16 … 400 Hz)
Max. load resistor
Ʋ
Operating voltage U
B
= 12 V DC
400 Ω
Ʋ
Operating voltage U
B
= 15 V DC
600 Ω
Electrical protective measures
Potential separation
Electronics potential free up to 500 V AC
Protection rating
Connection technology
Protection according to
EN 60529/IEC 529
Protection according to UL 50
M12 x 1 plug
IP66/IP67/IP69
Type 6P
Altitude above sea level
5000 m (16404 ft)
Protection class
III
Pollution degree
4
13.2 IO-Link
Technology
Overview
IO-Link is a technology standardized worldwide according to IEC 61131-9 for the communication
of a sensor with an IO-Link master. Point-to-point communication is based on the three-wire sensor
with corresponding IO-Link interface. This interface enables the master to access measurement
data and diagnostic information as well as to parameterize the sensor in the operating state.
Each IO-Link device has an IODD (IO Device Description). This is a device description file, in which
manufacturer, article number, functionality etc. are contained.
You can find the IODD file on our homepage as well as on the IODD finder of the IO-Link community.
IO-Link - physical layer
IO-Link specification: Revision 1.1
SIO mode: Yes
Speed: COM2 38.4 kBaud
Min. cycle time 4.0 ms
Length process data word: 32 Bit
IO-Link Data Storage: Yes
Block parameter adjustment: Yes