42
39089-US-210509
Supplemental Information
- Life Expectancy
20 years
- Temperature
-
40 °C … + 60 °C (- 40 °F … + 140 °F)
- Pressure
Atmospheric
- Vibration
Mild (less than 5 g)
- Corrosion
Mild
- Fire and Explosion
Accident Conditions Only
- Humidity
0% to 100% Relative Humidity
Licensing
A general license for the use of this source holder
must meet the requirements as outlined in OAC 3701:1-
46-05 (10 CFR 31.5 equivalent). Any other use of the
source holder requires a specific license.
Testing
- Leak Test
At least once every 36 months
- Shutter Test
At least once every 6 months
Transfer and Disposal
The transfer of this source holder containing the sealed source
is only to a specifically licensed person or as specified in OAC
3701:1-46 (an equivalent to 10 CFR 31 and 32). If you need
to transfer the source holder to another location or dispose of
the source holder, specific licensees are required.
Handling, Installation,
and Storage
A general licensee may initially mount the device. You may not
put the source holder into service until it is inspected by a
specific licensee. The manufacturer, or other specifically
licensed persons, are the only people that may perform service.
Storage
Whenever you remove the device from its mounted
position, mount the cover plate on the device to shield
the beam port and keep foreign material out of the
shutter assembly.
Physical Inspection
Perform physical inspections of the device and its
labeling at least once every six months for corrosion
prevention and maintenance in accordance with the
manufacturer’s instructions.
Summary of Contents for SHLG
Page 1: ...Operating Instructions SHLG Source Holder Document ID 39089 Radiometric...
Page 2: ......
Page 4: ...4 39089 US 210509 Revision Table NOTES...
Page 14: ...14 39089 US 210509 For Your Safety NOTES...
Page 32: ...32 39089 US 210509 Operation NOTES...
Page 40: ...40 39089 US 210509 Dismounting NOTES...
Page 47: ......