39089-US-210509
31
Operation
5.3 Special Conditions
Radiation Fields Inside Vessels
You must measure the radiation fields inside a vessel when entering the vessel.
Use the following statement when defining those requirements:
•
Equipment that is mounted on vessels or has accessible air gaps must
have written lockout procedures to ensure that access to the high levels
of the primary radiation beam is not possible.
•
In North America, many of these types of installationf are considered by
OSHA (Occupational Safety and Health Administration) as confined
spaces and the radiation is classified as “other known energy hazards”.
These installations require lockout documentation consistent with OSHA
rules and regulations.
•
You may use this documentation to meet the requirements of the U.S.
NRC or other regulatory agency.
At a minimum, follow the steps listed below, where applicable:
1.
Verify, by signature, that source holder is locked in the OFF (shutter
closed) position.
2.
Ensure that one person, who controls individuals entering the air gap,
holds the key.
3.
Make certain the radiation field is very low by using the gauge electronics
or a survey meter.
4.
Check that the gauge electronics shows a high reading or maximum
density, since no radiation is reaching the detector.
5.
Make sure the survey meter allows for quantitative measurements. The fields
measured should match very closely to background readings.
6.
Ensure that the vessel is vacant and the entrance door is locked, after the
work is completed.
7.
Complete these steps before inserting the source holder and returning it to
the ON position.
Summary of Contents for SHLG
Page 1: ...Operating Instructions SHLG Source Holder Document ID 39089 Radiometric...
Page 2: ......
Page 4: ...4 39089 US 210509 Revision Table NOTES...
Page 14: ...14 39089 US 210509 For Your Safety NOTES...
Page 32: ...32 39089 US 210509 Operation NOTES...
Page 40: ...40 39089 US 210509 Dismounting NOTES...
Page 47: ......