Weekly
➢
Check cooling flow, check for air flow by using “wrist test”. With the system powered on face
the machine and hold your exposed wrist in front of the air outlet on the left hand side. If no
airflow is felt inspect the inlet filter located on the right. Using your hands gently brush any
visible contamination away and check for airflow. If this does not correct air flow replace the
filter. If airflow is still not felt call service.
➢
General Clean Up, gently dry dust the inner deck and tablet (both closed and open) with a
tissue to remove visible dust and contamination
Monthly
➢
Change the air filters, using factory supplied replacement filters, with the system off remove
the filter frame and dispose of the old filter. Put the new filter in the frame and snap into
position. ALWAYS DISCONNECT THE SYSTEM FROM THE MAIN ELECTRICAL
OUTLET BEFORE SERVICING.
➢
Check posts for tightness; verify that the fiber and hand piece management pins are tight.
Gently wiggle each pin, if it feels lose tighten with a Philips screw driver, careful not too tight
a quarter turn past snug is sufficient.
➢
Visually inspect the case externals making sure the latches and handles are not lose from
attachment or damaged. If so contact service.
➢
Check functionality of the Emergency Switch and Remote Interlock.
Annually
➢
Factory Recertification, these systems require factory test and inspection on an annual basis.
Contact service to arrange for shipping back to the factory. This recertification is required to
maintain all extended service plans.
On the instruction of Sound field service system fuse change is performed as follows:
The system fuses are located in a holder in the upper part of the plug power module. Never attempt to
service the fuses when the plug is connected. To replace use two new BK/S500-3.15-R Glass fuse
3.15A 250VAC 5X20MM.
1.
Remove the fuse holder by inserting a small screw driver or similar into the slot located in the
center as shown. Don’t use a sharp object like a knife that could damage the fuse holder.
2.
Pull firmly directly away from the system to remove the holder. This does take a bit of force.
3.
Remove both fuses by first tilting the fuse in the holder as shown before removing.
4.
Change both fuses at the same time.
5.
Reverse this procedure to reinstall be sure that the holder is flush with the receptacle.