UM0000001 Rev_0 (2019/11/25)
Page
30
of
36
Copyright 2019 Sound Technologies
5810 Van Allen Way, Carlsbad CA 92008
800.268.5354 | soundvet.com
CAUTION!
USE EXTREME CAUTION WHEN DEVIATING FROM MANUFACTURER
RECOMMENDED SETTINGS.
CAUTION!
USE OF CONTROLS OR ADJUSTMENTS OR PERFORMANCE OF
PROCEDURES OTHER THAN THOSE SPECIFIED HEREIN MAY RESULT IN HAZARDOUS
RADIATION EXPOSURE.
8.6.3
Performing Treatment
Once all the parameters have selected the
Tissue Type
to be treated, the
Locality
of the problem, its
Chronicity
, the patients
Skin Color
and patients
Size,
you will be able to perform the treatment. The
chronicity of the treatment will dictate the number of phases for the treatment. Acute treatments will
have only 1 phase, while chronic treatments will have 2 phases.
Note
– Do not press the hand-switch until you see “Status: Ready” on the Treatment control, as this
safety measure will cause an error, this is to ensure a user is in positive control of the hand switch.
Release the hand-switch at the end of each phase and do not press again until you see “Status Ready”
on the treatment control again.
To get started, ensure that the
Begin Treatment
button shows Status: Standby. If there are any
problems with the communication between the program and the laser, it will read Status:
Communication Error. Once you see that it shows a standby status, tap the Begin Treatment button.
You will then see a message telling you to put your safety glasses on as shown below:
Tap this button one more time and you will hear a countdown in a woman’s voice and when the
system is ready for treatment the main button will appear as follows:.
CAUTION!
- Be sure to keep the laser in motion during the procedure, as opposed to holding it in
one place on the patient.