43
5.2.
X-Ray generator specifications
5.2.1.
X-Ray Tube Specification
(D-051)
z
Tube voltage:
50 ~ 100 kV
z
Tube current:
1 ~ 22 mA
z
Focal
spot:
0.5
mm
z
Inherent
filtration:
0.8
mm
Al
z
Added
filtration:
2.0
mm
Al
z
Total
filtration:
2.8
mm
Al
z
Filament characteristics:
3.5~4.9V 3.5A(max. filament current)
z
Anode
angle:
5°
z
Anode
Hu:
28000J
z
Anode cooling rate:
265W
z
Input energy at 1 sec:
1750W
z
Tube target material:
Tungsten