Installation manual
23
2. Please take out the wires, ONE72A, ONE74A and ONE75A, from the Cephalometric unit through
the guide hole on the vertical frame unit when
A
and
B
are close enough each other.
But leave wire ONE73A out from this work, which will be connected to the LAN Hub later.
ONE72A, ONE74A and ONE75A are
concerned with connection this time
3. Upon combining
A
part with
B,
fasten these two units using 2 wrench bolts (M8*60+8
Φ
flat
washer/spring]: Accessory No.15) and 1 wrench bolt(M8*50+8
Φ
flat washer/spring]:Accessory
No.16) at rear column.
4. Please fix the Column top cover (Accessory No.02) onto the back side of column, using wrench
bolts (M3*8 and M4*8)(Accessory No. 17 and 18)
M8*50
M8*60
M3*8
M4*8
Summary of Contents for PaX-Reve3D
Page 1: ......
Page 2: ......
Page 9: ...Installation manual 7 1 2 General view of the PaX Reve3D...
Page 41: ...Installation manual 39 5 1 4 Dimension of beam limiting device...
Page 42: ...PaX Reve3D 40 5 1 5 Dimension of the Unit Unit mm...
Page 54: ......
Page 55: ......