3. Acquiring PANO Images
PaX-i Plus User Manual
19
EN
G
LISH
3.2.1 PANO Examination Mode (Standard / Right / Left / Front /
Orthogonal)
Normal Patient Positioning
1.
Insert the Normal Bite into the Normal Chinrest and cover it with a Sanitary Vinyl
Cover.
The Sanitary Vinyl Cover is for single use only. It should
be replaced for each patient. Be sure to use the approved
vinyl cover.
Clean the Chinrest and the Bite with ethanol and wipe with
a dry towel before the next patient.
2.
Loosen the
Temple Supports OPEN/CLOSE Wheel
under the control panel to
widen the Temple Supports.
3.
Guide the patient to the inside of the equipment.
4.
Use the
Column UP/DOWN
button or switch option to adjust the height of the
equipment so that the patient's chin reaches the Chinrest.