4. Acquiring CEPH Images (Optional)
PaX-i Plus User Manual
47
EN
G
LISH
5.
Use the
Column UP/DOWN
button or switch option to adjust the height of the CEPH
unit to approximately match the height of the patient.
After adjusting the height of the column, align the Ear Rods
to the patient.
6.
During the operation, properly align the Ear Rods to the patient's ears so his/her
head does not move.
7.
Align horizontally so the patient's Frankfurt Line is parallel with the floor.
8.
Direct the patient to swallow first before closing his/her mouth and to remain in
his/her current position until image acquisition is completed.
9.
Click the
READY
button on Console Software. No X-ray will be emitted at this point.
10.
Now go
to 4.3 X-ray Exposure
to start the exposure.