3. Acquiring PANO Images
26
PaX-i Plus User Manual
3.2.2 SPECIAL Examination Mode (TMJ / Sinus)
<TMJ Open Mode (LAT / PA)>
The TMJ Close image can be acquired after the TMJ Open image is acquired
.
Steps for TMJ Mode
Patient positioning for TMJ Open > Laser Beam Aligning > X-
ray Exposure > Patient positioning for TMJ Close > Laser
Beam Aligning > X-ray Exposure
Patient Positioning
1.
Remove any existing Chinrest if it’s inserted and insert the
TMJ/Sinus Bite
.
Clean the Chinrest with ethanol and wipe with a dry towel
before the next patient.
2.
Loosen the
Temple Supports OPEN/CLOSE Wheel
under the control panel to
widen the Temple Supports.
3.
Guide the patient to the inside of the equipment.