9. Device Specifications
74
EzRay Air Portable
(Model: VEX-P300) User Manual
9.3
Electrical Specifications
Item
Description
Tube Voltage
Option 1: 60 kV
fixed (± 5 %)
Option 2: 65 kV
fixed (± 5 %)
Tube Current
2.5 mA (± 10 %)
Exposure Time
0.05-1.0 s (± 3 % or 10 ms)
Rated Voltage
21.6 V
dc
The system will be available with a fixed tube voltage
specification based on the user selection.
Maximum exposures on a full charge: about 600 times at 65
kV, 2.5 mA, 0.2 s
9.4
Environmental Specifications
Item
Description
During operating
Temperature
10 ~ 35
℃
Relative humidity
30 ~ 75 %
Atmospheric pressure
860 ~ 1060 hPa
Transport and storage
Temperature
-10 ~ 60
℃
Relative humidity
10 ~ 75 % non-condensing
Atmospheric pressure
860 ~ 1060 hPa
Failure to follow the specifications above can result in severe
injury and device damage.