Page 14
© 2013 Cascade Designs, Inc.
Installation Instructions
Seat Depth Adjustment
Seat depth adjustment varies according to the clinical application and the recline angle
of the back. In most cases, Icon Back System hardware allows seat depth to be adjusted
so that the back does not take up seat depth in front of the canes. Icon Back System
hardware also may be installed anterior of the canes in situations where seat depth is
desired for growth. Note: the Icon Back System can be mounted in a rotated position using
the VariLock Slide Braket Kit, ordered seperately. This kit will allow the left or right Slide
Brackets to be mounted in the anterior or posterior position, independently of one another.
Cane Bracket:
Anterior
Slide Bracket:
Anterior
Trunk Support Surface Range:
1.8 in. anterior of the canes to 3.3
in. anterior of the canes
Measured from front of canes to center of the Shell.
Cushion thickness ranges from 1.25 in to 2.75 in. depending on cushion inflation.
From the BASIC installation, remove the Holsters from the Cane Brackets. Swap the left and
right Cane Brackets and mount them on the wheelchair canes in the anterior position. Mount
the Holsters to the Cane Brackets ensuring that the Slide Caps open rearwards.
EXTRA GROWTH
Cane Bracket:
Posterior
Slide Bracket:
Anterior
Trunk Support Surface Range:
2.4 in. posterior of the canes to 0.9
in. posterior of the canes
Measured from front of canes to center of the Shell.
Cushion thickness ranges from 1.25 in to 2.75 in. depending on cushion inflation.
BASIC
Cane Bracket:
Posterior
Slide Bracket:
Posterior
Trunk Support Surface Range:
3.9 in. posterior of the canes to 2.4
in. posterior of the canes
Measured from front of canes to center of the Shell.
Cushion thickness ranges from 1.25 in to 2.75 in. depending on cushion inflation.
From the BASIC installation, remove the left and right Slide Brackets from the Shell Brackets.
Swap the left and right Slide Brackets and reattach them in the posterior position with the
tabs facing up.
EXTRA DEPTH
Cane Bracket:
Anterior
Slide Bracket:
Posterior
Trunk Support Surface Range:
0.3 in. anterior of the canes to 1.8
in. anterior of the canes
Measured from front of canes to center of the Shell.
Cushion thickness ranges from 1.25 in to 2.75 in. depending on cushion inflation.
From the BASIC installation, remove the Holsters from the Cane Brackets. Swap the left and
right Cane Brackets and mount them on the wheelchair canes in the anterior position. Next,
remove the left and right Slide Brackets from the Shell Brackets. Swap the left and right
Slide Brackets and mount them in the posterior position with the tabs facing up. Mount the
Holsters to the Cane Brackets ensuring that the Slide Caps open rearwards.
GROWTH
Installation Type Configuration