INSTRUCTIONS FOR USE
45
87-900-855-01 (B)
GENERAL INFORMATION
This equipment is destined for use by professionals.
The user should read this instruction manual and any
other additional information supplied by Varian before
operating the equipment. Varian will not be held
responsible for any events occurring due to non-
compliance, even partial, with these instructions,
improper use by untrained people, non-authorized
interference with the equipment or any action contrary
to that provided for by specific national standards. The
Turbo-V 6000 series controllers are microprocessor-
controlled, solid-state, frequency converters with self-
diagnostic and self-protection features.
The controllers drive (within ten steps) the
Turbo-V 6000 by the pump. They incorporate all the
facilities required for the automatic operation of the
Turbo-V 6000 pump series.
Remote start/stop, pump status signals, forepump
start/stop, interlock control (for pressure switch, water
flow switch. etc.) capability, are provided via auxiliary
connectors. The following paragraphs contain all the
information necessary to guarantee the safety of the
operator when using the equipment. Detailed
information is supplied in the appendix "Technical
Information".
This manual uses the following standard protocol:
!
The warning messages are for attracting the attention
of the operator to a particular procedure or practice
which, if not followed correctly, could lead to serious
injury.
CAUTION!
The caution messages are displayed before
procedures which, if not followed, could cause damage
to the equipment.
NOTE
The notes contain important information taken from the
text.
STORAGE
When transporting and storing the controllers, the
following environmental requirements should be
satisfied:
•
temperature: from -20 °C to + 70 °C
•
relative humidity: 0 - 95%
(without condensation)
PREPARATION FOR INSTALLATION
The controller is supplied in a special protective
packing. If this shows signs of damage which may have
occurred during transport, contact your local sales
office. When unpacking the controller ensure that it is
not dropped or subjected to any form of impact. Do not
dispose of the packing materials in an unauthorized
manner. The material is 100% recyclable and complies
with EEC Directive 85/399.
Controller packing
Each controller is factory set for a specific power
supply:
•
model 969-9491 is factory set for 220 Vac
operation
•
model 969-9591 is factory set for 120 Vac
operation
Check voltage selector window for correct set and
connect power cord.
WARNING!