INSTRUCTIONS FOR USE

45

87-900-855-01 (B)

GENERAL INFORMATION

This equipment is destined for use by professionals.
The user should read this instruction manual and any
other additional information supplied by Varian before
operating the equipment. Varian will not be held
responsible for any events occurring due to non-
compliance, even partial, with these instructions,
improper use by untrained people, non-authorized
interference with the equipment or any action contrary
to that provided for by specific national standards. The
Turbo-V 6000 series controllers are microprocessor-
controlled, solid-state, frequency converters with self-
diagnostic and self-protection features.
The controllers drive (within ten steps) the
Turbo-V 6000 by the pump. They incorporate all the
facilities required for the automatic operation of the
Turbo-V 6000 pump series.
Remote start/stop, pump status signals, forepump
start/stop, interlock control (for pressure switch, water
flow switch. etc.) capability, are provided via auxiliary
connectors. The following paragraphs contain all the
information necessary to guarantee the safety of the
operator when using the equipment. Detailed
information is supplied in the appendix "Technical
Information".

This manual uses the following standard protocol:

!

The warning messages are for attracting the attention
of the operator to a particular procedure or practice
which, if not followed correctly, could lead to serious
injury.

CAUTION!

The caution messages are displayed before
procedures which, if not followed, could cause damage
to the equipment.

NOTE

The notes contain important information taken from the
text.

STORAGE

When transporting and storing the controllers, the
following environmental requirements should be
satisfied:

  temperature: from -20 °C to + 70 °C

  relative humidity: 0 - 95%

(without condensation)

PREPARATION FOR INSTALLATION

The controller is supplied in a special protective
packing. If this shows signs of damage which may have
occurred during transport, contact your local sales
office. When unpacking the controller ensure that it is
not dropped or subjected to any form of impact. Do not
dispose of the packing materials in an unauthorized
manner. The material is 100% recyclable and complies
with EEC Directive 85/399.

Controller packing

Each controller is factory set for a specific power
supply:

  model 969-9491 is factory set for 220 Vac

operation

  model 969-9591 is factory set for 120 Vac

operation

Check voltage selector window for correct set and
connect power cord.

WARNING!

Summary of Contents for Turbo-V 6000

Page 1: ...NOTICE DE MODE D EMPLOI MANUAL DE INSTRUCCIONES MANUAL DE ISTRU ES BEDRIJFSHANDLEIDING ISTRUKSTIONSBOG BRUKSANVISNING INSTRUKSJON MANUAL OHJEK SIKIRJA INSTRUCTION MANUAL Turbo V 6000 controller Model...

Page 2: ...Turbo V 6000 Controller...

Page 3: ......

Page 4: ......

Page 5: ...ERCONNECTIONS 52 Connection P1 Logic Input Interconnections 53 Connection J2 Logic Output Interconnections 54 Controller to Pump Connection 55 Connection P31 Turbo V Pump Signals 55 Connection J21 Acc...

Page 6: ...tion 63 Monitor Relay Programming 66 Speed Adjustment 70 Starting the Pump 70 Operating the Pump 71 Low Speed Operation 71 Pump Shutdown 72 Power Failure 72 Remote Control Mode Operation 72 RS 232 Con...

Page 7: ...ono riportate tutte le informazioni necessarie a garantire la sicurezza dell operatore durante L utilizzo dell apparecchiatura Informazioni dettagliate sono fornite nell appendice Technical Informatio...

Page 8: ...a sia appoggiata su di un tavolo assicurarsi che sia stabile Non fare funzionare mai la pompa se la flangia di ingresso non collegata al sistema o non chiusa con la flangia di chiusura NOTA Il connett...

Page 9: ...guasto Prima di effettuare qualsiasi intervento sul controller scollegare il cavo di alimentazione Qualora un controller dovesse essere rottamato procedere alla sua eliminazione nel rispetto delle nor...

Page 10: ...la temperatura sul radiatore dei Mosfets di uscita superiore a 60 C Attendere che la temperatura ritorni al di sotto della soglia Premere due volte il pulsante START per riavviare la pompa FAULT OVERL...

Page 11: ...Bedieners bei der Anwendung des Ger ts aufgef hrt Detaillierte technische Informationen sind im Anhang Technical Information enthalten In dieser Gebrauchsanleitung werden Sicher heitshinweise folgend...

Page 12: ...t die Pumpe auf einem Tisch mu auf den stabilen Stand geachtet werden da sonst die Gefahr von Personen und Ger tesch den besteht Die Pumpe nie einschalten wenn der Eingangsflansch nicht am System ange...

Page 13: ...der Netzstecker gezogen werden Eine eventuelle Verschrottung mu unter Einhaltung der einschl gigen landes blichen Vorschriften erfolgen FEHLERMELDUNGEN In einigen St rungsf llen zeigt das Selbstdiagno...

Page 14: ...trollers hat 90 C berschritten oder die Temperatur am Radiator der Mosfets betr gt ber 60 C Warten bis die Temperatur unter den Schwellenwert gesunken ist Die Pumpe durch zweimalige Bet tigung der STA...

Page 15: ...ndications n cessaires garantir la s curit de l op rateur pendant l utilisation de l appareillage Des renseignements plus d taill s se trouvent dans l appendice Technical Infor mations Cette notice ut...

Page 16: ...te derni re est stable Ne jamais faire fonctionner la pompe si la bride d entr e n est pas connect e au sy st me ou si elle n est pas ferm e l aide de la bride de serrage NOTE Laisser le connecteur de...

Page 17: ...e op ration sur le contr leur d brancher le c ble d alimentation En cas de mise au rebut d un contr leur proc der son limination conform ment aux r glementations natio nales en la mati re MESSAGES D E...

Page 18: ...adiateur des Mo sfets de sortie est sup rieure 60 C Attendre que la temp rature retourne au dessous du seuil Presser deux fois l interrupteur START pour remettre la pompe en marche FAULT OVERLOAD Pend...

Page 19: ...necesaria para garantizar la seguridad del operador durante el uso del equipo Una informaci n m s detallada se facilita en el Suplemento Technical Information Este manual utiliza los s mbolos convenci...

Page 20: ...esa cerciorarse que es estable No poner en marcha nunca la bomba si la brida de entrada no est conectada al sistema o no est cerrada con la brida de cierre NOTA El conector di cierre J1 ha de dejarse...

Page 21: ...do Antes de efectuar cualquier operaci n en el controler desenchufar el cable de alimentaci n En caso de que un controler se tenga que desguazar efectuar su eliminaci n respetando las normas nacionale...

Page 22: ...el radiator de los Mosfets de salida excede los 60 C Esperar a que la temperatura vuelva por debajo del umbral Apretar dos veces el pulsador START para volver a poner en marcha la bomba FAULT OVERLOA...

Page 23: ...eguintes est o descritas todas as informa es necess rias para garantir a seguran a do operador durante o uso da aparelhagem Informa es pormenorizadas s o fornecidas no ap ndice Technical Information E...

Page 24: ...apoiada numa mesa certificar se que esteja est vel Nunca activar a bomba se o flange de entrada n o estiver ligado ao sistema ou n o estiver fechado com o flange de fecho NOTA O conector de fecho J1...

Page 25: ...er com defeito Antes de efectuar qualquer opera o no controller desligar o cabo de alimenta o Caso um controller deva ser destru do proceder sua elimina o respeitando as normas nacionais espec ficas M...

Page 26: ...ler superou os 90 C ou a temperatura no radiador dos Mosfets de sa da superior a 60 C Aguardar at que a temperatura volte ao limite estabelecido Premer duas vezes o bot o START para reactivar a bomba...

Page 27: ...akelaars van de waterstroom enz In de volgende paragrafen is alle informatie vermeld om de veiligheid van de operator tijdens het gebruik van de apparatuur te verzekeren Gedetailleerde informatie is t...

Page 28: ...Laat de pomp nooit werken zonder dat de ingangsflens aan het systeem is gekoppeld of de afsluitflens is gesloten OPMERKING De connector J1 moet met zijn jumper aangesloten blijven als geen externe aan...

Page 29: ...er ter vervanging van de defecte controller Alvorens werkzaamheden aan de controller uit te voeren de voedingskabel afkoppelen Mocht de controller gesloopt worden ga dan overeenkomstig de specifieke n...

Page 30: ...verschreden of de temperatuur op de radiator van de uitgangsMosfets bedraagt meer dan 60 C Wacht tot de temperatuur weer onder de drempelwaarde is gezakt Bedien twee maal de START knop om de pomp weer...

Page 31: ...lder al information der beh ves for at garantere operat rens sikkerhed under anvendelsen Detaljeret information findes i bilaget Technical Information I brugsanvisningen anvendes f lgende standar drub...

Page 32: ...mpen st r fast hvis den er installeret p et bord Dette er for at forebygge skader p apparatet og personer Start aldrig pumpen hvis pumpetill bet ikke er tilsluttet systemet eller er blokeret BEM RK Af...

Page 33: ...den der er g et i stykker Inden der foretages noget som helst indgreb p styreenheden skal str mmen f rst afbrydes Skrotning af pumpen skal foreg i overensstemmelse med det p g ldende lands s rlige lov...

Page 34: ...r temperaturen p radiatoren til Mosfet kredsl bene i udgangen er over 60 C Vent p at temperaturen falder til under t rskelv rdi Tryk to gange p START for at starte pumpen igen FAULT OVERLOAD Under nor...

Page 35: ...all information som beh vs f r att garantera operat rens s kerhet under driften Detaljerade uppgifter finns i bilagan Technical information I bruksanvisningen anv nds f ljande standar drubriker Varnin...

Page 36: ...tta f r att f rebygga skador p apparaten och personer S tt aldrig ig ng pumpen om intagsfl nsen varken r kopplad till systemet eller r blockerad p plats med l sfl nsen OBSERVERA St ngningskontakten J1...

Page 37: ...renhet Innan n got arbete utf rs p styrenheten m ste dess str mf rs rjning brytas Skrotning av pumpen skall ske enligt g llande lagstiftning FELMEDDELANDEN N r vissa fel uppst r visar styrenhetens sj...

Page 38: ...formator har verstigit 90 C eller temperaturen vid kylaren till Mosfet utg ngskretsar verstiger 60 C V nta tills temperaturen sjunker under tr skelv rdet Tryck tv g nger p knappen START f r att starta...

Page 39: ...all informasjon som er n dvendig for sikre brukeren n r utstyret er i bruk For mer detaljert bruk vises det til tillegget Technical Information Denne veiledningen bruker f lgende standard protokoll D...

Page 40: ...ig for forhindre skader p apparatet og p personer Dersom inngangsflensen hverken er tilkoplet systemet eller dersom den er blokkert av l seflensen m pumpen aldri startes opp MERK Lukkekontakten J1 m v...

Page 41: ...yreenheten Disse F r noe arbeid utf res p styreenheten m den frakoples str mnettet Dersom en styreenhet skal kasseres m dette skje i henhold til nasjonale bestemmelser FEILMELDINGER N r det oppst r vi...

Page 42: ...or har overskredet 90 C eller temperaturen p kj leren til Mosfet kretsl pene i utgangen er over 60 C Vent til temperaturen synker under terskelverdien Trykk to ganger p knappen START for starte pumpen...

Page 43: ...illa on luettavissa tarpeelliset tiedot laitteen k ytt j n turvallisuuden takaamiseksi laitteen k yt n aikana Yksityiskohtaiset tiedot l ytyv t liitteest Tekniset tiedot T m k sikirja k ytt seuraavia...

Page 44: ...istakaa ett se on vakaa N in v ltyt n vammoilta ihmisille sek itse koneelle lk my sk n k ytt k pumppua mik li sis ntulon laippaa ei ole kytketty j rjestelm n tai mik li sit ei ole suljettu laippasulki...

Page 45: ...en mink tahansa valvojaan teht v n toimenpiteen suorittamista irroittakaa s hk kaapeli pistorasiasta Mik li valvoja t ytyy romuttaa toimikaa sen h vitt misess kansallisten s d sten ja normien m r v ll...

Page 46: ...uuntajan l mp tila on ylitt nyt 90 C tai ulostulon Mosfets piirikorttien j hdyttimen l mp tila on yli 60 C Odottakaa ett l mp tila putoaa kynnysarvon alapuolelle Painakaa painonappia START kaksi kerta...

Page 47: ...le oi aparaivthte plhroforive pou egguouvntai thn asfavleia tou ceiristhv katav th diavrkeia th crhsimopoivhsh th suskeuhv Leptomereiv plhroforive parevcontai sto paravrthma Technical Information Auto...

Page 48: ...ovti eivnai kalav staqeropoihmevnh Mhn qevtete se leitourgiva thn antliva an h flavntza eisovdou den eivnai sundedemevnh sto suvsthma hv an den eivnai kleisthv me thn flavntza kleisivmato SHMEIWSH O s...

Page 49: ...mevnou Prin kavnete opoiadhvpote epevmbash ston Ruqmisthv aposundevste to kalwvdio trofodovthsh Gia thn katastrofhv tou ruqmisthv akolouqhvste ovti anafevrete stou eqnikouv kanonismouv MHNUMATA LAQOUS...

Page 50: ...ivnai anwvterh twn 60 C Perimevnete wvspou h qermokrasiva na katebeiv kavtw apov to anwvtato epitreptov shmeivo Pathvste duvo forev to koumpiv START gia na qevsete xanav se kivnhsh thn antliva FAULT O...

Page 51: ...ailed information is supplied in the appendix Technical Information This manual uses the following standard protocol The warning messages are for attracting the attention of the operator to a particul...

Page 52: ...p is laying on a table make sure it is steady Never operate the Turbo V pump if the pump inlet is not connected to the system or blanked off NOTE The input signal J1connector should be left in positio...

Page 53: ...ers are available on an advance exchange basis through Varian Before carrying out any work on the controller disconnect it from the supply If a pump is to be scrapped it must be disposed off in accord...

Page 54: ...perature exceeds 90 C or the temperature on the outgoing Mosfets radiator exceeds 60 C Wait until the temperature decrease below threshold value Press the START push button twice to start the pump FAU...

Page 55: ...uding power supply and 3 phase output Control PCB including analog and input output section microprocessor and digital section The power supply converts the single phase 50 60 Hz AC mains supply into...

Page 56: ...og output 0 10 V J21 power output interc 120 Vac disregarding the mains 3A for heater band 1 A for forepump relay coil and Vent Device P31 signal inputs Temperature sensors Speed detector Oil level de...

Page 57: ...TECHNICAL INFORMATION 51 87 900 855 01 B CONTROLLER OUTLINE The outline dimensions for the Turbo V 6000 controllers are shown in the following figures Controller models 969 9491 and 969 9591 outline...

Page 58: ...TECHNICAL INFORMATION 52 87 900 855 01 B INTERCONNECTIONS The following figure shows the Controller interconnections Controller models 969 9491 and 969 9591 interconnection...

Page 59: ...anently closed contact relay contact transistor etc When the contact closes the turbopump runs at low speed and when the contact opens the turbopump reverts to high speed mode With the remote mode ope...

Page 60: ...y isolated output pin 11 positive pin 3 negative The output Voltage will be present when the START push button on front panel is pressed or the remote start is present or the function has been request...

Page 61: ...tion P31 connector Signals cable Connection J21 Accessories and Options Interconnections The controller is provided with a power connector plug J21 Ensure that the main circuit breaker is off then rem...

Page 62: ...n or equivalent 9 pin D type male connector The external cable not supplied between the host computer and the controller does not require crossed wires so that signals are connected correctly For exam...

Page 63: ...of the value contained in a window the slave answers a string equal to the one sent by the master but in addition there is the field DATA containing the value of the window The format of the field DA...

Page 64: ...Destination PC 02 80 32 30 30 30 30 30 30 2E 30 30 03 39 44 STX ADD WINDOW RD 000 00 ETX CRC Command FREQUENCY Source PC Destination Inverter 02 80 32 30 33 30 03 38 32 STX ADDR WINDOW RD ETX CRC Sou...

Page 65: ...In the case of reading request of a window the data field doesn t exist ETX 0x03 CRC XOR among all the characters following STX with exception of STX including the end character ETX hexadecimally enc...

Page 66: ...S NO 121 X X L HEATER BAND ENABLE YES NO 200 X N CURRENT A 201 X N VOLTAGE V 202 X N POWER W 203 X N EXCITEMENT FREQUENCY Hz 205 X N PUMP STATE 0 6 STOP WAITING INTERLOCK STARTING NORMAL HIGH LOAD FAI...

Page 67: ...Start mode allows the pump to ramp up to Normal Speed slowly with a minimum ramp up time of 80 minutes and a maximum of about 105 minutes The Soft Start mode is always operative as default mode If it...

Page 68: ...X X C Y Y X X X K R P M X X X X W where UP Upper bearing temperature LO Lower bearing temperature MOT Motor temperature XX X KRPM Rotation speed XXXX W Power absorption by the controller 0 to 2 000 W...

Page 69: ...conds and the processor enters in a routine where it is possible to program the controller In this routine the CYCLE push button is used for choosing changing the value or condition the PUMP CURRENT p...

Page 70: ...ct YES Press PUMP CURRENT push button and the display shows E X T E R N A L I N P U T N O where NO is the default value Press the CYCLE NUMBER key to select YES Press PUMP CURRENT push button and the...

Page 71: ...te for the host computer or printer communication The controller is factory set for 9600 baud rate operation Instead of RS 232 one of configured RS 232 RS 485 RS 422 ports will be displayed Select the...

Page 72: ...pump If PRINT is selected and a printer is connected on RS 232 line an unidirectional communication is established and every minute the data are sent to the printer even if the pump is not running Th...

Page 73: ...ld value b pump speed threshold value after run up time c pump speed threshold value before run up time programmed speed threshold value R1 ON OFF a b c R1 operation R2 ON OFF a b c R2 operation delay...

Page 74: ...g the PUMP CURRENT push button and the display shows R U N U P T I M E S E L X X h X X m X X s where RUN UP TIME is the interval time from start to speed threshold value in hours minutes seconds Selec...

Page 75: ...ld value see the preceding cycle diagram This function is factory set to YES D E L A Y E V E N A F T E R T H R E S H O L D Y E S Press PUMP CURRENT to confirm and the display shows P U M P L I F E X X...

Page 76: ...MBER and PUMP CURRENT push buttons are always active even when the operating mode selected is REMOTE or RS 232 422 485 Press the START push button or use the remote or serial port start signal and the...

Page 77: ...to reach normal operation The pump can be stopped at any rotational speed and can be restarted at any rotational speed from either the front panel buttons or the remote connections The controller aut...

Page 78: ...nly if the power failure is longer than the preset delay time When power is restored the Turbo V controller automatically restarts the interconnected devices and the turbopump in the proper sequence T...

Page 79: ...on kit 969 9858 RS 485 computer communication kit 969 9856 P2 output mating connector 969 9852 J1 input mating connector 969 9853 Mains cable European plug 3m long SR 03 660441 01 Mains cable American...

Page 80: ...TECHNICAL INFORMATION 74 87 900 855 01 B...

Page 81: ...TIFICATION Product Description Varian P N Varian S N Purchase Reference TYPE OF RETURN check appropriate box Paid Exchange Paid Repair Warranty Exchange Warranty Repair Loaner Return Credit Shipping E...

Page 82: ...rate No zero high backround Gauge tube not working Display problem Vacuum system unstable Cannot reach test mode Communication failure Degas not working Failed to start Other Error code on display Oth...

Page 83: ...an India PVT LTD 101 108 1st Floor 1010 Competent House 7 Nangal Raya Business Centre New Delhi 110 046 India Tel 91 11 5548444 Fax 91 11 5548445 Italy Varian Vacuum Technologies Via F lli Varian 54 1...

Reviews: