TECHNICAL INFORMATION
62
87-900-855-01 (B)
•
Press the PUMP CURRENT push-button and the
display shows:
U P
X X º C
L O = X X º C
M O T = X X º C
Y Y
X X . X
K R P M
X X . X
A
where:
•
UP
= Upper bearing temperature
•
LO
= Lower bearing temperature
•
MOT
= Motor temperature
•
XX.X KRPM
= Rotation speed
•
XX.X A
= DC absorption by the controller
00.0 to 25.0 A
•
YY
= during operation a selected set point
condition (1 or 2 contrast inverted) appears
when the programmed threshold speed
value is not reached.
•
Press the PUMP CURRENT push-button and the
display shows:
U P
X X º C
L O = X X º C
M O T = X X º C
Y Y
X X . X K R P M
X X X X
W
where:
•
UP
= Upper bearing temperature
•
LO
= Lower bearing temperature
•
MOT
= Motor temperature
•
XX.X KRPM
= Rotation speed
•
XXXX W
= Power absorption by the
controller 0 to 2.000 W
•
YY
= during operation a selected set point
condition (1 or 2 contrast inverted) appears
when the programmed threshold speed
value is not reached.
•
Press the PUMP CURRENT push-button and the
display shows:
U P
X X º C
L O = X X º C
M O T = X X º C
Y Y
X X . X K R P M
X X X
H z
where:
•
UP
= Upper bearing temperature
•
LO
= Lower bearing temperature
•
MOT
= Motor temperature
•
XX.X KRPM
= Rotation speed
•
XXX Hz
= frequency range 12 to
234 Hz
•
YY
= during operation a selected set point
condition (1 or 2 contrast inverted) appears
when the programmed threshold speed
value is not reached.
Summary of Contents for Turbo-V 6000
Page 2: ...Turbo V 6000 Controller...
Page 3: ......
Page 4: ......