TECHNICAL INFORMATION
7
4 87-900-940-01 (
F
)
Controller-to-Pump Connection
A five-meter long cable is provided to connect the
controller to the pump. The following figures show the
controller output connector configuration where pins:
•
A-F = pump temperature sensor
•
B-C-D = 75 Vac 3-phase output to pump motor
stator
•
E = ground
Controller-to-Pump Connector
RS 232 OPTION
RS 232 Installation
To install it, proceed as follows:
•
Switch off the power and disconnect the power
cord.
•
Unscrew the cover screws and remove the cover.
•
On rear panel, remove the J5 connector plate.
•
Insert the flat cable through the rear panel and plug
it into the socket located on the internal board of the
controller.
•
Install and tighten the connector screws.
•
Replace the cover.
RS 232 connector installation
RS 232 Communication Port
The communication serial port connections and the
minimum connection configuration are shown in the
following figures.
Communication Serial Port Connections
Minimum Connection Configuration
The communication port mating connector is supplied
with the RS 232 PCB (AMP/Cannon or equivalent 15-
pin "D" type male connector).
The external cable (not supplied) between the host
computer and the controller doesn’t require any
crossed wires so that the signal are connected
correctly.
For example, the Transmit data signal from controller
(pin 2) must be connected to the host computer's
Receive data line (pin 2) and vice versa. Consult the
host computer's instruction manual for its serial port
connections.
NOTE
Varian cannot guarantee compliance with FCC
regulations for radiated emissions unless all external
wiring is shielded, with the shield being terminated to
the metal shroud on the 0-subconnector. The cable
should be secured to the connector with screws.
PUMP
TEMPERATURE
SENSOR
75 V
3 Phase