TECHNICAL INFORMATION

69

87-900-852-01 (B)

With normal LOW SPEED operation, the pump will run
at about 2/3 of its nominal speed and achieves a base
pressure somewhat higher than the standard
specifications. If the gas load becomes higher, the
controller output frequency and voltage start to
decrease automatically, and the Turbo-V pump is
slowed down in proportion to the gas load until it
reaches 3 KRPM.
If the LOW SPEED mode is selected after normal
operating condition is reached, the display shows:

A P P R O A C H

I

N G

L S

X X

K R P M

L S

while approaching the low speed value.

When the low speed mode is deselected, the pump
starts to accelerate to its rotational speed. The display
shows:

N O R M A L

O P E R A T

I

O N

X X

K R P M

Pump Shutdown

Press the front panel STOP pushbutton or remove the
remote signal; the power from the turbopump will be
removed and the pump will begin to slow down.

Power Failure

In the event of a power failure (momentary or long term), the
Turbo-V controller will stop the turbopump and all the
interconnected pumps/devices. The Turbo-V vent valve
device, if used, will vent the turbopump only if the power failure
is longer than the preset delay time. When power is restored,
the Turbo-V controller automatically restarts the interconnected
devices and the turbopump in the proper sequence. The
display shows:

P U M P

I

S

S T A R T

I

N G

1

2

X X

K R P M

until normal operation achieved.

Remote Control Mode Operation

If remote signals are used to operate the controller, it
must be programmed for remote operation (see
paragraph "FRONT/REMOTE/232 Selection") and
when ready to start, the display shows:

R E A D Y

F O R

R E M O T E

S O F

T

S T A R T

If the Soft Start has been deselected the display shows:

P U M P

R E A D Y

:

U S E

R E M O T E

S T A R T

With or wihout Soft Start mode selected the
START/STOP and LOW SPEED front panel
pushbuttons are inoperative, while the CYCLE
NUMBER and PUMP CURRENT pushbuttons are
always active.

RS 232 Control Mode Operation

If the RS 232 option is installed and the controller has
been programmed for RS 232 operation, the controller
may be driven by a computer and when ready to
operate, the display shows:

R E A D Y

F O R

R S

2

3

2

S O F

T

S T A R T

If the Soft Start has been deselected the display shows:

P U M P

R E A D Y

:

U S E

R S

2

3

2

L

I

N E

With or wihout Soft Start mode selected the
START/STOP, LOW SPEED functions are under
computer control, while the CYCLE NUMBER and
PUMP CURRENT front panel pushbuttons are always
active.

Summary of Contents for Turbo-V 250 series

Page 1: ...CH NOTICE DE MODE D EMPLOI MANUAL DE INSTRUCCIONES MANUAL DE ISTRU ES BEDRIJFSHANDLEIDING ISTRUKSTIONSBOG BRUKSANVISNING INSTRUKSJON MANUAL OHJEK SIKIRJA INSTRUCTION MANUAL Turbo V 250 Controllers Mod...

Page 2: ......

Page 3: ...Turbo V 250 Controllers...

Page 4: ......

Page 5: ......

Page 6: ...NICAL INFORMATION 49 TURBO V 250 CONTROLLER DESCRIPTION 49 CONTROLLER SPECIFICATIONS 50 CONTROLLER OUTLINE 50 INTERCONNECTIONS 52 Connection P1 Logic Input Interconnections 53 Connection J2 Logic Outp...

Page 7: ...ction 61 Monitor Relay Programming 64 Starting the Pump 67 Operating the Pump 68 Low Speed Operation 68 Pump Shutdown 69 Power Failure 69 Remote Control Mode Operation 69 RS 232 Control Mode Operation...

Page 8: ...attirano l attenzione dell operatore su una procedura o una pratica specifica che se non eseguita in modo corretto potrebbe provocare gravi lesioni personali ATTENZIONE I messaggi di attenzione sono...

Page 9: ...rocedure operative Per ulteriori dettagli e per procedure che coinvolgono collegamenti o particolari opzionali fare riferimento al paragrafo Use dell appendice Technical Information Alcune delle istru...

Page 10: ...ualsiasi intervento deve essere eseguito da personale autorizzato 1 Connettore di ingresso dei segnali logici il connettore di accoppiamento viene fornito con l apposito ponticello di richiusura 2 Con...

Page 11: ...ro 15 minuti X pu valere da 0 a 9 ed indica il passo non superato Verificare che il sistema non presenti delle perdite Premere due volte il pulsante START per riavviare la pompa FAULT PUMP OVERTEMP La...

Page 12: ...Die Gefahrenhinweise lenken die Aufmerksamkeit des Bedieners auf eine bestimmte Prozedur oder Praktik die bei unkorrekter Ausf hrung schwere Verletzungen hervorrufen k nnen ACHTUNG Die Warnhinweise vo...

Page 13: ...Betriebsvorg nge aufgef hrt F r weitere Hinweise bez glich Anschl sse und Montage des bestellbaren Zubeh rs siehe Kapitel Use im Anhang zu Technical Information Einige der nachstehenden Gebrauchshinw...

Page 14: ...V 250 sind wartungsfrei Eventuell erforderliche Eingriffe m ssen von dazu befugtem Fachpersonal ausgef hrt werden 1 Eingangsverbinder der logischen Signale der Kupplungs verbinder wird mit einer Wied...

Page 15: ...cht X kann 0 9 sein und zeigt den nicht berwundenen Schritt an Sicherstellen da das System keine Leckagen aufweist Die Pumpe durch zweimalige Bet tigung der START Taste neustarten FAULT PUMP OVERTEMP...

Page 16: ...Les messages de danger attirent l attention de l op rateur sur une proc dure ou une manoeuvre sp ciale qui si elle n est pas effectu e correctement risque de provoquer de graves l sions ATTENTION Les...

Page 17: ...tous autres d tails et pour les proc dures concernant des connexions ou des l ments en option se reporter au paragraphe Use de l appendice Technical Information Quelques unes des instructions donn es...

Page 18: ...n exigent aucun entretien Toute op ration doit tre effectu e par un personnel agr 1 Connecteur d entr e des signaux logiques le connecteur d enclenchement est dot de la barrette de r enclenchement sp...

Page 19: ...oir de 0 9 et indique le pas non franchi S assurer que le syst me ne pr sente pas de fuites Presser deux fois l interrupteur START pour remettre la pompe en marche FAULT PUMP OVERTEMP La temp rature d...

Page 20: ...obre un procedimiento o una operaci n espec fica que al no realizarse correctamente podr a provocar graves lesiones personales ATENCI N Los mensajes de atenci n se visualizan antes de procedimientos q...

Page 21: ...m s detalles y para procedimientos que impliquen conexiones u opcionales especiales les remitimos al apartado Use del anexo Technical Information Algunas de las instrucciones citadas a continuaci n p...

Page 22: ...er operaci n ha de ser efectuada por personal autorizado 1 Conector de entrada de las se ales l gicas el conector de acoplamiento se suministra con el conector puente espec fico de cierre 2 Conector d...

Page 23: ...erado Comprobar que el sistema no presenta p rdidas Apretar dos veces el pulsador START para volver a poner en marcha la bomba FAULT PUMP OVERTEMP La temperatura del rodamiento superior o de la bomba...

Page 24: ...ens de perigo chamam a aten o do operador para um procedimento ou uma pr tica espec fica que se n o efectuada correctamente pode provocar graves les es pessoais ATEN O As mensagens de aten o s o visua...

Page 25: ...es detalhes e para procedimentos que envolvem liga es ou pe as opcionais consultar o par grafo Use do ap ndice Technical Information Algumas das instru es descritas a seguir s o poss veis no controlle...

Page 26: ...nuten o Todas as opera es devem ser efectuadas por pessoal autorizado 1 Conector de entrada dos sinais l gicos o conector de acoplamento fornecido com ponte espec fica de fecho 2 Conector de sa da dos...

Page 27: ...vista a cada passo em 15 minutos X pode valer de 0 a 9 e indica o passo n o superado Verificar se o sistema n o apresenta perdas Premer duas vezes o bot o START para reactivar a bomba FAULT PUMP OVERT...

Page 28: ...Deze handleiding hanteert de volgende symbolen Bij dit symbool staat tekst die de aandacht van de operator vestigt op een speciale procedure of methode die indien niet correct uitgevoerd ernstig licha...

Page 29: ...treffen wordt verwezen naar de paragraaf Use van de bijlage Technical Information Enkele hier vermelde instructies zijn alleen van toepassing op het controllermodel 969 9425 indien hierop het accessoi...

Page 30: ...o V 250 zijn onderhoudsvrij Eventuele werkzaamheden moeten door bevoegd personeel worden uitgevoerd 1 Connector voor ingang logische signalen de te koppelen connector wordt met de speciale sluitklem g...

Page 31: ...aarde tussen 0 en 9 aannemen en geeft de niet overschreden stap aan Controleer of het systeem geen lekkage vertoont Bedien twee maal de START knop om de pomp weer op te starten FAULT PUMP OVERTEMP De...

Page 32: ...rselsmeddelelserne informerer operat ren om at en speciel procedure eller en vis type arbejde skal udf res pr cist efter anvisningerne I modsat fald er der risiko for sv re personskader VIGTIGT Denne...

Page 33: ...ret beskrivelse samt procedurer der involverer tilslutninger eller tilbeh r henvises til afsnittet Use i bilag Technical Information Visse procedurer der beskrives nedenst ende kan kun udf res p styre...

Page 34: ...else Ethvert indgreb p pumpen skal foretages af autoriseret personale 1 Indgangskontakt til logiske signaler tilslutningskontakten leveres med specielt forbindelsesbro 2 Udgangskontakten til logiske s...

Page 35: ...ekte hastighed for hvert trin indenfor 15 minutter X kan g lde 0 til 9 og angiver hvilket trin der ikke blev gennemf rt Kontroll r om systemet er ut t Tryk to gange p START for at starte pumpen igen F...

Page 36: ...na informerar operat ren om att en speciell procedur eller en viss typ av arbete m ste utf ras exakt enligt anvisningarna I annat fall finns risk f r sv ra personskador VIKTIGT Detta varningsmeddeland...

Page 37: ...samt betr ffande moment som involverar anslutningar eller tillbeh r h nvisas till avsnittet Use i bilaga Technical Information Vissa anvisningar som listas nedan kan endast utf ras p styrenheten model...

Page 38: ...h llsfria Allt servicearbete m ste utf ras av auktoriserad personal 1 Ing ngskontakt f r logiska signaler anslutningskontakten levereras med speciell st ngningsbrygga 2 Utg ngskontakt f r logiska sig...

Page 39: ...nuter X kan g lla 0 till 9 och anger vilket steg som inte genomf rts Kontrollera om systemet l cker Tryck tv g nger p knappen START f r att starta om pumpen FAULT PUMP OVERTEMP Temperaturen p det vre...

Page 40: ...isse meldingene skal tiltrekke seg brukerens oppmerksomhet til en spesiell fremgangsm te eller praksis som hvis den ikke f lges kan medf re alvorlige skader FORSIKTIG Denne advarselen vises foran frem...

Page 41: ...som omfatter tilkoplinger eller ekstrautstyr vises det til avsnittet Use i vedlegget Technical Information Visse anvisninger som listes nedenfor kan kun gjennomf res p styreenheten modell 969 9425 n...

Page 42: ...d p styreenheten m kun utf res av autorisert personell 1 Inngangskontakt for logiske signaler tilkoplingskontakten leveres med spesiell lukkebrygge 2 Utgangskontakt for logiske signaler og kontroll av...

Page 43: ...orrekte hastigheten for hvert trinn i l pet av 15 minutter X kan gjelde 0 til 9 og angir hvilket trinn som ikke har blitt gjennomf rt Kontroller om systemet lekker Trykk to ganger p knappen START for...

Page 44: ...j n huomion kiinnittym n erityisiin toimintotatapoihin joiden seuraamatta j tt minen voi aiheuttaa vakavia henkil vaurioita HUOMIO Huomio merkinn t varoittavat toiminnoista joiden laiminly nti voi joh...

Page 45: ...Tarkempia lis tietoja sek kytkent j ett valinnaisia lis laitteita koskevien toimenpiteiden suorittamista k sitt vi tietoja l yd tte kappaleesta K ytt joka on Tekniset tiedot kappaleen liitteen Jotkin...

Page 46: ...valvojaan teht v t toimenpiteet tulee j tt aina valtuutetun henkil n teht viksi 1 Logiikkasegnaalien tulokytkent pariliittimet toimitetaan niihin sopivilla yhdyskaapelisulkijalla 2 Logiikkasegnaalien...

Page 47: ...e joka askeleella tarkoitettua nopeutta 15 minuutin sis ll X on askelnumero 09 ja ilmaisee sen askeleen jota ei ole ylitetty Tarkistakaa ett j rjestelm ss ei ole vuotoja Painakaa painonappia START kak...

Page 48: ...eiv ti akovlouqe sumbavsei Oi endeivxei kinduvnou proselkuvoun thn prosochv tou ceiristhv se mia diadikasiva hv se mia eidikhv ergasiva h opoiva eavn den ektelesteiv swstav qa mporouvse na prokalevsei...

Page 49: ...vreie kai gia diadikasive pou apaitouvn idiaivtere sundevsei hv axesouavr anaferqeivte sthn paravgrafo Crhvsh tou Pararthvmato Technical Information Merikev apov ti anaferovmene sth sunevceia plhrofor...

Page 50: ...hvpote epevmbash qa prevpei na pragmatopoihqeiv apov exousiodothmevno proswpikov 1 Buvsma eisovdou twvn logikwvn shmavtwn to buvsma zeugarwvmato promhqeuvetai me to eidikov gefuravki kleisivmato 2 Buv...

Page 51: ...cei xeperasteiv Elevgcete ovti to suvsthma den parousiavzei diarroev Pathvste duvo forev to koumpiv START gia na qevsete xanav se kivnhsh thn Antliva FAULT PUMP OVERTEMP H qermokrasiva tou avnw kouzin...

Page 52: ...e operator to a particular procedure or practice which if not followed correctly could lead to serious injury CAUTION The caution messages are displayed before procedures which if not followed could c...

Page 53: ...that involve optional connections or options are supplied in the paragraph USE of the appendix Technical Information Some procedures can be executed with controller 969 9425 only when the accessory H...

Page 54: ...terminal MAINTENANCE The Turbo V 250 series controller does not require any maintenance Any work performed on the controller must be carried out by authorized personnel 1 Logic input signals connector...

Page 55: ...anned value within 15 minutes X is the step number from 0 to 9 indicating the step number not passed Verify that system has no leaks Press the START pushbutton twice to start the pump FAULT PUMP OVERT...

Page 56: ...er transformer PCB including power supply and 3 phase output analog and input output section microprocessor and digital section display and keyboard circuits The power supply converts the single phase...

Page 57: ...ng the mains P1 optically insulated input Minimum ON 3 mA Maximum 5 mA J2 optically insulated output 24 Vdc 60 mA P6 output fan 100 mA J7 output forepump coil etc 120 Vac 1A disregarding the mains Rad...

Page 58: ...TECHNICAL INFORMATION 51 87 900 852 01 B Controller model 969 9425 outline...

Page 59: ...L INFORMATION 52 87 900 852 01 B INTERCONNECTIONS The following figure shows the Controller interconnections Controller models 969 9423 and 969 9523 interconnections Controller model 969 9425 intercon...

Page 60: ...anently closed contact relay contact transistor etc When the contact closes the turbopump runs at low speed and when the contact opens the turbopump reverts to high speed mode With the remote mode ope...

Page 61: ...V 60 mA optically isolated output pin 13 positive pin 6 negative The output voltage will be present when the START push button on front panel is pressed or the remote start is present or the function...

Page 62: ...cable are provided assembled with the air cooling kit Controller to Pump Connection A five meter long cable is provided or is available as an option for model 969 9425 to connect the controller to th...

Page 63: ...s 120 Vac with the maximum current of 1 A If more than one device is connected be sure to draw not more than 120 VA to avoid controller damage Use a relay coil with a maximum surge power of 65 VA NOTE...

Page 64: ...front panel from 600 to 9600 baud The controller is factory set for 9600 baud operation NOTE When a baud rate of 2400 4800 9600 baud is used the requested information from Host to controller must be...

Page 65: ...STARTING 3 NORMAL OPERATION 4 5 HIGH LOAD 6 FAILURE 7 APPROACHING LOW SPEED 2 5 Cycle Time 6 9 Pump Life 10 11 Pump Temperature 12 Current 13 Voltage 14 17 Frequency 18 19 Cycle 20 R1 Status 21 R2 St...

Page 66: ...3 NORMAL OPERATION 4 5 HIGH LOAD 6 FAILURE 7 APPROACHING LOW SPEED R2 status 0 OFF 1 ON R1 status 0 OFF 1 ON 2 CRC When the Request is J Numerical readings the Answer will contain a string of 5 charac...

Page 67: ...wer source The display lights and shows R E A D Y F O R L O C A L S O F T S T A R T The controller with the Soft Start mode allows the pump to ramp up to Normal speed slowly with a minimum ramp up tim...

Page 68: ...shbutton is used for choosing changing the value or condition the PUMP CURRENT pushbutton is used to enter and confirm the value At any time it is possible to exit this routine by pressing the CYCLE a...

Page 69: ...for FRONT panel operation Choose the desired selection by pressing the CYCLE pushbutton F R O N T R E M O T E 2 3 2 S E L E C T I O N F R O N T Press the PUMP CURRENT pushbutton to enter the value and...

Page 70: ...on Cycle time Pump life Pump temperature Pump current Pump voltage Controller output frequency Cycle number R1 condition R2 condition Pump life and cycle zeroing Configuration parameter readings Confi...

Page 71: ...RPM is the switch point of relay R1 at the preset turbopump speed adjustable from 00 to 99 KRPM The speed threshold will condition the R1 and R2 operation see the following cycle diagram and it is fac...

Page 72: ...the PUMP CURRENT pushbutton and the display shows R U N U P T I M E S E L X X h X X m X X s where RUN UP TIME is the interval time from start to speed threshold value in hours minutes seconds Select...

Page 73: ...d value see the preceeding cycle diagram This function is factory set to YES D E L A Y E V E N A F T E R T H R E S H O L D Y E S Press PUMP CURRENT to confirm and the display shows P U M P L I F E X X...

Page 74: ...M P I S S T A R T I N G 1 2 X X K R P M where 1 2 contrast inverted identifies the set point condition 1 is displayed when relay R1 is de energized and the related output is zero voltage 2 is display...

Page 75: ...the FRONT panel operation selected the remote and RS 232 operations are inoperative conversely the CYCLE NUMBER and PUMP CURRENT pushbuttons are always active even when the operating mode selected is...

Page 76: ...longer than the preset delay time When power is restored the Turbo V controller automatically restarts the interconnected devices and the turbopump in the proper sequence The display shows P U M P I S...

Page 77: ...9 9854 Mains cable European plug 3 m long SR 03 660441 01 Mains cable American plug 120 V 3 m long SR 03 660441 02 OPTIONS DESCRIPTION PART NUMBER RS 232 kit 969 9851 P2 output mating connector 969 98...

Page 78: ...TECHNICAL INFORMATION 71 87 900 852 01 B R...

Page 79: ...TECHNICAL INFORMATION 72 87 900 852 01 B...

Reviews: