TECHNICAL INFORMATION 

 19 

87-900-049-01 

(

E

)

 

OPERATION 

Make all vacuum manifold and electrical connections 
and refer to the ion pump instruction manual prior to 
operating the MiniVac controller. 
 

NOTE 

The accessory P001 connector should be left in 
position including the jumper between pin 3 and 7 if no 
external connection has been made. 

 
 

For small ion pumps, up to 8 l/s: 

with a roughing pump, establish a rough pressure 
suitable to start the ion pump. 
 

For medium ion pump (20 l/s to 60 l/s): 

establish a pressure lower than 10

-5

 mbar/ Torr. 

 

For large ion pumps (120 l/s to 400 l/s): 

establish a pressure lower than 2 x 10

-6

 mbar/Torr. 

 

 

  

CAUTION

 

Before plugging in the controller power cable, be sure 
that the selected operating voltage matches the power 
source to avoid equipment damage. 

 
 

• 

Plug the controller power cable into a suitable 
power source. 

• 

Place the ON/OFF switch in the ON position. 

• 

Monitor the VacIon pump current and the controller 
high voltage output by watching the LED lit, then 
read the corresponding current/voltage value (e.g.: 
if the third LED is increasing its brightness it 
means that the pump current is between 2 and 3 
mA and the high voltage output is between 5 and 4 
kV). 

• 

If only the first LED of the bar graph is lit, the 
current drawn by the pump is less than 1 mA and 
the voltage output is 5 kV. 

• 

If ten LEDs are lit, the current drawn by the pump 
is 10 mA and the voltage output is 2 kV. 

• 

For the correct determination of the voltage value, 
use the voltage vs current diagram shown in the 
figure “MiniVac Controller Rear Panel Controls”. 

 

Pressure Determination 

Pressure at the inlet flange of a VacIon pump is 
proportional to the current drawn by the pump. 
 
To determine the pressure of the inlet flange of the 
VacIon pump, determine the pump current displayed 
on the front panel or read the recorder output 
(especially for current lower than 1 mA), and find the 
corresponding pressure value from the current vs 
pressure curve of the relevant VacIon pump. 

Power Failure 

In the event of a power failure (momentary or long 
period) the controller is switched off. When power is 
restored, the controller will automatically restart. 

Remote Control Mode Operation 

It is possible to remotely switch onb and off the high 
voltage using a permanent contact connected to pins 3 
and 7 of J001 accessory connector with the mains 
power ON/OFF switch to ON. 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Summary of Contents for MiniVac 929-0190

Page 1: ...9 01 E APRIL 2006 I MANUALE DI ISTRUZIONI D BEDIENUNGSHANDBUCH F NOTICE DE MODE D EMPLOI GB INSTRUCTION MANUAL MiniVac Controller Model 929 0190 Model 929 0290 Model 929 0191 Model 929 0291 vacuum tec...

Page 2: ...MiniVac Controller...

Page 3: ......

Page 4: ......

Page 5: ...TIONS 15 CONTROLLER OUTLINE 16 INSTALLATION 17 Line Voltage Change Over 17 Output Polarity Selection 17 Input Output Interconnections 18 Pump Connection 18 OPERATION 19 Pressure Determination 19 Power...

Page 6: ...li ATTENZIONE I messaggi di attenzione sono visualizzati prima di procedure che se non osservate potrebbero causare danni all apparecchiatura NOTA Le note contengono informazioni importanti estrapolat...

Page 7: ...95 non condensante Per il collegamento del controller con la relativa pompa utilizzare il cavo specifico del controller stesso Per gli altri collegamenti e l installazione degli accessori opzionali ve...

Page 8: ...ura fornito con l unit ha i pin 3 e 7 cortocircuitati Remote ON OFF Per poter accendere l HV necessario montare questo connettore se non connesso un contatto di Remote Interlock NOTA L unit settata in...

Page 9: ...nnen ACHTUNG Die Die Warnhinweise vor bestimmten Prozeduren machen den Bediener darauf aufmerksam da bei Nichteinhaltung Sch den an der Anlage entstehen k nnen ANMERKUNG Die Anmerkungen enthalten wic...

Page 10: ...emperatur von 0 C bis 45 C Relative Luftfeuchtigkeit 0 95 nicht kondensierend F r den Anschlu des Controllers an die Pumpe mu das zum Controller geh rende Kabel benutzt werden Andere Anschl sse und di...

Page 11: ...ie Stifte 3 und 7 kurzgeschlossen Remote ON OFF Falls kein Remote Interlock Kontakt angeschlossen ist kann die Hochspannung nur angeschaltet werden wenn dieser Stecker montiert ist ANMERKUNG Fabriksei...

Page 12: ...ssent avant certaines proc dures qui si elles ne sont pas observ es pourraient endommager s rieusement l appareillage NOTE Les notes contiennent des renseignements importants isol s du texte EMMAGASIN...

Page 13: ...ve 0 95 non condensante Pour la connexion du contr leur la pompe correspondante utiliser le c ble du contr leur pr vu cet effet Pour les autres connexions et pour l installation des accessoires en opt...

Page 14: ...t a les pins 3 et 7 en court circuit Remote ON OFF Pour pouvoir allumer l HV il faut monter ce connecteur si un contact de Remote Interlock n est pas branch NOTE L unit est assign e l usine avec polar...

Page 15: ...correctly could lead to serious injury CAUTION The caution messages are displayed before procedures which if not followed could cause damage to the equipment NOTE The notes contain important informati...

Page 16: ...specific cable supplied with the controller See the section Technical Information for detailed information about the above mentioned and the other connections and about the options installation USE T...

Page 17: ...are short circuited Remote ON OFF This loopback plug must be fitted to power on the HV unless a Remote Interlock contact is connected NOTE The unit is factory set with a negative output polarity Refe...

Page 18: ...be operated at any pressure up to 2 x 10 6 mbar Torr continuous operation Medium larger ion pumps must first be started with an appropriate control unit and pumped to a sufficiently low pressure where...

Page 19: ...ut scale kV 4 Controller power ON OFF switch 5 Current voltage LED s bar graph indicator 6 H V ON LED 7 Recorder output connector 8 Pump current scale mA The MiniVac controller rear panel controls are...

Page 20: ...nector 0 to 10 Vdc 10 V correspond to 1 mA resolution of 10 mV minimum corresponding to 1 A Available at pins 2 and 6 of rear panel Accessory connector Voltage Recorder Output Rec Out 0 to 5 V corresp...

Page 21: ...TECHNICAL INFORMATION 16 87 900 049 01 E Voltage Vs Current Diagram CONTROLLER OUTLINE The outline dimensions of the MiniVac controller are shown in the following figure MiniVac Outline Dimensions...

Page 22: ...el check back door for voltage selector set Using a small screwdriver pull out the voltage selector and fuses Select the operating voltage then firmly insert the voltage selector and fuses in place Ch...

Page 23: ...pin 2 signal out pin 6 ground The output voltage 0 to 10 Vdc is proportional to the pump current 0 to 1 mA e g 10 Vdc correspond to 1 mA The resolution is 10 mV minimum corresponding to 1 A Pin 4 8 Vo...

Page 24: ...ness it means that the pump current is between 2 and 3 mA and the high voltage output is between 5 and 4 kV If only the first LED of the bar graph is lit the current drawn by the pump is less than 1 m...

Page 25: ...e lit Switch power off and disconnect the high voltage cable Switch power on and verify that only one LED is lit If this happens the high voltage cable or the pump are shorted If not replace the unit...

Page 26: ...TIFICATION Product Description Varian P N Varian S N Purchase Reference TYPE OF RETURN check appropriate box Paid Exchange Paid Repair Warranty Exchange Warranty Repair Loaner Return Credit Shipping E...

Page 27: ...rate No zero high backround Gauge tube not working Display problem Vacuum system unstable Cannot reach test mode Communication failure Degas not working Failed to start Other Error code on display Oth...

Page 28: ...1263 Toll Free 0120 655 040 Korea Varian Technologies Korea Ltd Shinsa 2nd Bldg 2F 966 5 Daechi dong Kangnam gu Seoul Korea 135 280 Tel 82 2 3452 2452 Fax 82 2 3452 2451 Toll Free 080 222 2452 Mexico...

Reviews: