TECHNICAL INFORMATION
15
87-900-049-01
(
E
)
CONTROLLER SPECIFICATIONS
Operation
All type and size of
VacIon pump
Input:
- Voltage
(4 ranges)
- Frequency
- Power
- Fuse
100 Vac
120 Vac
220 Vac
240 Vac
50 to 60 Hz
50 VA
2 x T 2A
(disregarding the mains)
Output:
- Voltage
- Current
- Power
±5000 Vdc ±10% (refer to
the following figure) (polarity
through internal card)
15 mA (short circuit) (refer to
the following figure)
21 W
Dual Current
Recorder
Output:
- Rec Out1
- Rec Out2
0 to 10 Vdc (10 V correspond
to 10 mA);
resolution of 10 mV minimum
(corresponding to 10 µA)
Available on front panel BNC
connector and at pins 1 and 6
of rear panel Accessory
connector.
0 to 10 Vdc (10 V correspond
to 1 mA);
resolution of 10 mV
minimum (corresponding
to 1 µA)
Available at pins 2 and 6 of
rear panel Accessory
connector.
Voltage
Recorder
Output:
- Rec Out
0 to 5 V (corresponding to
0 to 5 kV with 1 V
corresponding to 1 kV in
linear mode)
Available at pins 4 and 8 of
rear panel Accessory
connector.
Operating
temperature
0 °C to 45 °C
Storage
temperature
-20 °C to +70 °C
Radio interference
suppression
EN 55011 class A Group 1
EN 61000-4-2
EN 61000-4-4
EN 61000-4-3
Safety EN
61010-1
Installation category II
Pollution degree
2
Output connector:
- High voltage
- Accessory
FISCHER type:
part number 105 A049
for model 929-0290 and
929-0191
KINGS type:
part number 1064-1
for model 929-0190 and 929-
0291
J 001 input/output
signal (9-pin "D" type socket)
Cable
Mains, 3 meters long
Weight