Testing Specifications
Specification Table
210-MS
Ionization
Mode
Test Sample
Amount
Measured
m/z RMS
S/N
Specification
Library Match
(any search type)
EI
PN 393065201
(1 pg/µL OFN, 5 pg/µL BZP)
1 pg OFN
272
≥
50:1
≥
500
220-MS
Ionization
Mode
Test Sample
Amount
Measured
m/z RMS
S/N
Specification
Library Match
(any search type)
EI
PN 393065201
(1 pg/µL OFN, 5 pg/µL BZP)
1 pg OFN
272
≥
50:1
≥
500
CI
PN 39065201
(1 pg/µL OFN, 5 pg/uL BZP)
5 pg BZP
183
≥
20:1
N/A
EI Methodology
NOTE: The MS comes with a CD containing factory test method and data files.
Use these method files to pass the installation sensitivity tests. At the factory, the
diluted OFN standard is injected using a pressure pulse and splitless injection
technique. Retain the Test Files CD for your records.
1. Ensure the MS temperatures are set as follows:
•
Trap: 150 °C
•
Manifold: 40 °C
•
Transfer Line: 230 °C
2. Click
Activate Method
from the
MS Workstation
menu.
3. Open
the
OFN Method
in the
Service Folder
.
The following shows
System Control
in a
Ready
state.
System Control
is
ready when both the GC and the MS are ready.