Do not place the foreline pump on the bench with the MS. Vibrations from
the foreline pumps will degrade the MS performance.
4. Place the pump under the bench.
Do not lengthen the vacuum hoses to prevent degrading the vacuum.
The foreline pumps must be connected to a ventilation system.
NOTE: You may observe foam when you look at the oil level. The oil may foam
until sufficient oil collects on the filter to create a seal and then it will stop
foaming.
03-954143-55:Rev. 1