User Manual: UM19893-EN
Harmony3 series
Revision: 5.0
Date of release: 2019-01-29
13
/20
www.vareximaging.com
4.
Level the detector and lock it behind the grip(s) in the front.
WARNING:
When inserting the detector, make sure the detector is detected correctly. If the detector
is inserted correctly the
Detector Present and correctly inserted
LED lights up (Optional
feature)
Removing the detector
1.
Open the drawer until it locks in its end position.
2.
Push the detector backwards and lift it over the front grip(s).
3.
Remove the detector from the drawer.
4.
Closing the drawer
1.
Push the knob next to the handle to unlock the end position of the drawer.
WARNING
Always use both hands while closing the drawer: one hand to push the knob next to the
handle and one hand to hold the handle.
2.
Push the drawer back inside the Harmony3 series.
The LED’s will indicate the position of the drawer (portrait or landscape) and if the image
receptor is present or not. (this is an optional feature)
WARNING
Make sure the detector is at its end position before closing the drawer. Closing the drawer
when the detector is not fully inserted can damage the detector and/or Harmony3 series.
Figure 1: Incorrectly inserted drawer. The
image receptor is not inserted completely
Figure 2: Correctly inserted drawer. The
image receptor is at the right position