Specifications and accessories
7.1.
Technical specifications
Description
Reference
Classifications
See § 2.10
Application
Holding an image receptor, part of an X-ray system
Materials
ABS, POM, stainless steel, aluminium
Power source (ST and NR version
excluded)
12-42 [VDC] ± 10% / 24 [VA]
Detector/Grid present contact
(Harmony3 series and NR version)
Contact closed when detector/grid is present.
Max. 42VDC @ 2A
Grid detection contact
(Harmony3 series ST version)
Contact closed when grid inserted.
Max. 42VDC @ 2A
Weight
±23 kg (Harmony3 series)
±12 kg (ST version only)
Al eq.
0.23[mm] @ 75[kV] (table models)
0.32[mm] @ 75[kV] (Wall models)
Size
Table: 571x584x69mm (WxDxH)
Wall: 610x610x77mm (WxDxH)
Operation environment
Ambient temperature: 10 – 40 °C
Relative humidity:
10 – 75 %, non-condensing
Atm. Pressure:
690 – 1080 hPa
Storage
Ambient temperature: -10 – +70 °C
Relative humidity:
10 – 90 %, non-condensing
Atm. Pressure:
690 – 1080 hPa
Applicable standards
IEC 60601-1
IEC 60601-1-2
IEC 60601-1-3
IEC 60601-2-54