Part
Procedure
Comments
Replace the dry filter into the inlet opening
making sure that the filter is completely tucked
into the opening and that there are no gaps
around the filter.
This procedure will help to assure the product’s
performance and service life by allowing the
maximum amount of clean, dust-free air to be
available for cooling the compressor motor.
Replace the filter every 4000 hours or if there
are any signs of tears, degradation or clogging
of the filter material.
Water Trap
and Air Hose
Exterior
Wipe the exterior
according to hospital
disinfection policies.
Use clean water to
remove any residual
residue.
Do not allow liquid
to enter the outlet
of the water trap as
this liquid will enter
the Precision Flow
device causing
damage to the
electrical and/or
mechanical
components.
Water Bottle
Reservoir
Wipe the exterior
of the water bottle
reservoir and
reservoir tubing
according to hospital
disinfection policies
after emptying the
water bottle reservoir
and in between
patient uses. Rinse
the water bottle with
warm soapy water in
between patients.
Confirm rinsing of the
water bottle reservoir
to remove soap
residue.
Page 23
3100772 Rev E