Starting the Compressor

After connecting the power cord, toggle the power inlet switch into the 

“I ” position, after which the cooling fan, compressor, and hour meter 

will turn on simultaneously.

WARNING: Use only the power cord supplied with the 

compressor.

WARNING:  To  avoid  risk  of  electric  shock,  this  equipment 

must only be connected to a supply mains with 

protective earth.

WARNING: Ensure the air hose has been properly connected 

to the Precision Flow unit prior to powering on the 

compressor.

WARNING: Ensure the Precision Flow is powered off before 

starting the compressor.

After powering on the compressor, plug in the Precision Flow and 

begin therapy.

Running

Although the compressor itself utilizes fans to cool the motor and 

compressor sleeves, the Q50 has been designed with a secondary 

fan to provide additional cooling to the compressor as well as 

reduce the moisture content and temperature of the air being 

provided to the Precision Flow unit. The compressor has an internal

Page 19

3100772  Rev E

Summary of Contents for Q50

Page 1: ...Q50 Compressor Instructions for Use...

Page 2: ......

Page 3: ...and Weight 7 High Altitude Considerations 7 Working Principle and Construction 9 Working Principle 9 System Overview 9 Set Up for Use 13 Mounting Positions 13 Pole Mount 14 Install High Altitude High...

Page 4: ...oxygen supply to the DISS oxygen fitting on the oxygen filter assembly on the back of the Precision Flow if oxygen concentrations above 21 are desired and pressurize the oxygen source 7 Attach the Q50...

Page 5: ...lied air The compressor was designed to be mounted on a Precision Flow Roll Stand or placed directly on a flat level floor that does not obstruct the air exhaust underneath the compressor The low prof...

Page 6: ...ould be used ONLY by a trained operator Contraindications The Compressor is not intended for use in MRI environments For more information see the Precision Flow Instructions for Use Warnings and Cauti...

Page 7: ...de High Flow Kit Auto draining water trap assembly Water Bottle Water Bottle Holder Drain Tube 1 Pole mounts and one air intake filter are installed during manufacturing Related Documents Warranty Cer...

Page 8: ...IEC 60601 1 medical electrical equipment Part 1 general safety requirement Environmental Requirements Relative Temperature Humidity Operation 18 30 C 10 90 Storage 10 50 C 20 90 CAUTION If the storag...

Page 9: ...level conditions The density of air is lower at higher altitudes and has an impact on the flow capacity of the compressor The following table provides guidance for operating at higher altitudes CAUTI...

Page 10: ...1 31 40 40 40 40 40 39 37 34 32 32 40 40 40 40 40 39 37 34 32 33 40 40 40 40 40 40 38 35 33 34 40 40 40 40 40 40 38 35 33 35 40 40 40 40 40 40 38 36 34 36 40 40 40 40 40 40 39 36 34 37 40 40 40 40 40...

Page 11: ...he heat exchange coil and into the compressor Note Excess moisture is directed to the water bottle reservoir Here it is compressed moves through the heat exchange coil in order to reduce the air tempe...

Page 12: ...Intake Filter 2 Power Inlet Switch and Fuse Drawer 3 Hour Meter 4 Serial Number Label 5 DISS Air Fitting Figure 2 Air Intake Filter Power Inlet Switch and Fuse Drawer Hour Meter DISS Air Fitting Seria...

Page 13: ...rth Air Intake The compressor utilizes air from its surrounding environment It passes through multiple stages of filtration prior to delivery to the Precision Flow The first filtration stage is access...

Page 14: ...compressor This could result in failure to deliver therapy as intended When the unit is placed on the floor be sure that any flooring such as carpet does not restrict the air exhaust WARNING Do not re...

Page 15: ...compressor the correct distance from the floor to accommodate cooling air to exhaust from the unit WARNING Do not remove the compressor feet These are designed to reduce noise and provide the require...

Page 16: ...the compressor must be placed on the Precision Flow Roll Stand base and no greater than 8 from the ground The compressor also must be placed over two locking casters of the Precision Flow Roll Stand I...

Page 17: ...right side place the Vapotherm compressor on the right of the roll stand base Note The side with the water bottle reservoir should face front 2 PlaceonehalfofapoleclampbetweentheVapothermcompressor mo...

Page 18: ...this ring will release the fitting on the water trap Pull the water trap away from the Precision Flow 2 Install the drain tube into the fitting on the bottom of the water trap Gently pull on the tube...

Page 19: ...e end of the air hose onto the DISS air fitting on the back of the Q50 Compressor until it reaches the threads Rotate the collar on the air hose clockwise so that the threads engage with the DISS air...

Page 20: ...or and Precision Flow unit before turning the unit on 4 Connect oxygen supply to Precision Flow if applicable Connect an oxygen supply as directed in the Precision Flow Instructions for Use if oxygen...

Page 21: ...s been properly connected to the Precision Flow unit prior to powering on the compressor WARNING Ensure the Precision Flow is powered off before starting the compressor After powering on the compresso...

Page 22: ...drain tube into the cap of the water bottle and pressing the water bottle into the water bottle holder NOTE It is normal for some water to remain in the water trap bowl while draining the majority of...

Page 23: ...nic emissions IEC 61000 3 2 Not applicable The Vapotherm Q50 compressor is suitable for use in all establishments other than domestic and those directly connected to the public low voltage power suppl...

Page 24: ...water bottle reservoir and reservoir tubing according to hospital disinfection policies after emptying the water bottle reservoir and in between patient uses Do not allow liquids to infiltrate beyond...

Page 25: ...l Water Trap and Air Hose Exterior Wipe the exterior according to hospital disinfection policies Use clean water to remove any residual residue Do not allow liquid to enter the outlet of the water tra...

Page 26: ...g used CAUTION To prevent damage to the unit Always remove power cord prior to any cleaning or disinfection of the compressor If there are any questions regarding the safety or use of the disinfecting...

Page 27: ...r 4 000 hours Air intake muffler replacement once per 4 000 hours Compressor rebuild once per 8 000 hours Compressor motor replacement once per 20 000 hours Replace the filter element Vapotherm Part N...

Page 28: ...e Power cord is not connected or connected to a faulty outlet Connect power cord and ensure that the outlet is functioning Power cord is connected to a functioning power outlet but there is no audible...

Page 29: ...an is running but not the compressor even when the compressor is cool The compressor has failed Call for service The compressor and cooling fan are running The air pipeline is not tightened properly T...

Page 30: ...rt on bottom of compressor The fan has failed Call for service Air flow required of the compressor exceeds capabilities FiO2 and flow rates exceed the guidance for high altitude operation provided on...

Page 31: ...paid by the Customer Customer shall also be responsible for the cost of labor for repairs The warranty set forth herein shall become null and void if 1 the Q50 device is not used or serviced in accord...