3001002 Rev. T
Page 13
Oxygen
Air
Power
cord
Pole
clamp
Section 7
Setting up
7-1. Hang the sterile water supply from hook on
Vapotherm approved roll stand.
7-2. Attach the unit to Vapotherm approved roll stand
below lowest point of the sterile water supply.
NOTE
: The Precision Flow
®
oxygen and air supply
inlet fittings are gas-specific to ensure correct connection.
WARNING:
Unit weighs 10.6 lb.(4.81kg) To prevent
possible injury or damage from falling, it must be
securely fixed to a Vapotherm approved roll stand,
with the base of the unit not more than 40” (102cm)
above the floor. Fixed rail supports may also be used.
Use with Vapotherm approved roll stands.
7-3. Connect oxygen and air supply hoses to correct
inlets, then connect them to the wall outlets.
7-4. Connect power cord.
7-5. Open the bags containing the disposable water path,
vapor transfer cartridge, and delivery tube, and
assemble them as follows:
7-5-1. Remove the rubber plugs from the vapor transfer cartridge.
Install a high or low-flow vapor transfer cartridge in disposable waterpath as shown.
The vapor transfer cartridge may be inserted either way up. Align the vapor transfer
cartridge ports with the disposable water path openings and press firmly into place.
A High Flow Vapor Transfer Cartridge is shown below. It is indicated with a
REF: PF-VTC-HIGH and blue caps. High Flow Cartridges are for flow rates of 5-40 LPM.
A Low Flow Vapor Transfer Cartridge is shown below. It is indicated with a
REF: PF-VTC-LOW, red caps and the addition of two black stripes. Low Flow
Cartridges are for flow rates of 1-8 LPM.