3001002 Rev. T
Page 27
Section 16
Specifications
PERFORMANCE
Oxygen percentage:
Range- 21 to 100% O
2
Accuracy- ± 2%
Resolution- 1%
NOTE: At 22% & 23% oxygen blend, the delivered oxygen is 21%. At 98% and 99% oxygen
blend, the delivered oxygen is 100%.
Flow rate:
STANDARDS
Designed to conform to the following standards:
IEC 60601-1
UL60601-01
CSA C.22.2/No. 601.1
AS/NZS 3200.1.2
EN60601-1
ISO 8185
ISO 11195
ISTA-2A
ENVIRONMENTAL
Operating
Ambient temperature: 18-30°C
Ambient relative humidity: 20-90% RH non-condensing
Ambient Pressure: Standard atmospheric – Not to be used in hyperbaric conditions
Storage and Shipping
Ambient temperature: -10 - +50°C
Ambient relative humidity: 20-90% RH
ALARM SOUND PRESSURE RANGES
Medium Priority Alarm
47 dB measured 1m from unit
Low Priority Alarm
45 dB measured 1m from unit
EXPECTED LIFE
The Precision Flow
®
has an expected life of 5 years with typical use of roughly 150 days
per year. Actual life will vary depending on whether typical use is greater than 150 days
per year, quality of gas supplies, cleaning with proper agents, operating per Instructions
for Use instructions and cautions, and the timely execution of routine maintenance. More
frequent use, dirty and or wet gas supplies, or the use of organic or abrasive cleaners will
reduce the expected life.