REPLACE OXYGEN
SENSOR
Section 16 Specifications
PHYSICAL CHARACTERISTICS
Dimensions:
Height 11.5”(300mm), width 8”(200mm), depth 7”(180mm), excluding Vapotherm approved
roll stand clamp and gas inlet filters.
Weight:
10.6 lb (4.81 kg) without disposable patient circuit
Circulating Water Volume:
400 ml approx. including delivery tube and vapor transfer cartridge.
Mounting
:
Rear mounted clamp fits Vapotherm approved roll stands up to 1.5”(38mm) diameter.
Gas Connections
:
Standard non-interchangeable fittings for medical air and oxygen.
FUSES:
(Qty 2) GMA 3A F250 V 5mm x 20mm
SYSTEM REQUIREMENTS
Power:
100-240VAC, 50-60Hz, approx. 200VA during warm up, approx. 80VA in steady state
(depends on flow rate and temperature).
Back-up power:
4.8V nickel-metal hydride battery pack.
Gas supply:
Medical air and oxygen at inlet pressures between 4 and 85 psi (28-586 KPa).
NOTE:
the full range of flows and oxygen percentage is available only if both gases are
present at inlet pressures of at least 40 psi (276 kPa). Unit is calibrated at Vapotherm
using 100% O
2
Water:
Sterile water for inhalation in pre-filled sealed container.
PERFORMANCE
Temperature:
Range- 33 to 43°C at exit from the delivery tube, adjustable
Resolution- 1°C
Accuracy- ± 2°C
Warm up time:
± 2°C of 33°C set point < 5 minutes (at ambient 23°C).
NOTE: Time required to heat is dependent on the temperature set point, flow rate and
ambient temperature. Due to normal system variation, it is possible not all combinations of
temperature and flow rate may be achieved, in particular when set to the higher settings.
Humidification:
Complies with ISO8185-2007 Respiratory tract humidifiers for medical use, paragraph 101