P/N VT-108732-US, Manual revision: C
Sealing Head Operation
3-4
Figure 3-3, Repeating the Seal Process
Warning: Be sure to
push
the tubing through the sealing region to the next
position. Under
no
circumstances should the tubing or bag be separated by
pulling it while the sealing jaws are clamped. This action may cause an opening
in the tubing, which will subject the blood or plasma to non-sterile conditions
and the user to potentially hazardous fluids. In addition, the Sealing Head may
become damaged by arcing due to contamination of the sealing region.
Warning: If you observe arcing while making a seal, follow the instructions in
Chapter 5, Table 5-1, "Arcing or Bad Seals." Although a seal made when you
experience arcing may look acceptable, it may be inadequate for centrifugation.
Take precautions as if this is an inadequate seal.