P/N VT-108732-US, Manual revision: C
Cleaning the Sealing Head
4-
4
S
PLASH
G
UARD
R
EPLACEMENT
Note: The Sealing Head assembly comes with a clear plastic splash guard to
provide operator protection in the event of a tubing rupture during sealing. It is
made with a self-adhesive coating for easy attachment to the lever. Extra splash
guards are provided with each shipment and are available for purchase from
Vante (order part number 10800061).
Warning: To avoid coming in contact with any blood or blood products, wear
disposable protective gloves while handling and cleaning the Sealing Head
components. See Chapter 6.
1.
Disconnect the Sealing Head from the Power Source by removing the RF
power cable and seven pin connector from the Power Source.
Warning: Failure to separate the Sealing Head from the power source prior to
removing or attaching the splash guard may result in an RF burn.
2.
Remove the existing splash guard from the Sealing Head and properly dis-
pose of it.
Warning: Always dispose of any materials that have come in contact with blood
or blood products in accordance with procedures provided in Chapter 6.
Figure 4-4, Removing Splash Guard from Sealing Head
3.
Remove the adhesive residue from the surface with a commercially avail-
able solvent cleaner such as De-Solv-It
®
, Goo Gone
®
or WD-40
®
, applied
to a cotton swab (see Figure 4-5, Removing Adhesive Residue).
Summary of Contents for SEBRA 2605
Page 1: ...operation manual VT 108732 US Model 2605 Segmenting Head VANTE INC IGNITING INNOVATION...
Page 2: ...Printed in USA Copyright 2002 2010 2017 Vante...
Page 4: ......
Page 7: ......
Page 12: ...P N VT 108732 US Manual revision C Sealing Head Description 2 4...
Page 18: ......
Page 30: ......
Page 31: ......