Vante

® 

Model 4600 

16 

3.2

 

Troubleshooting 

The following chart offers diagnosis and actions for many commonly-reported problems.  
For any problems or failures not listed, please contact the Vante Customer Service 
Department. 

Problem 

Diagnosis 

Action 

Not sealing. 

No power to Power 
Source. 

Ensure all connections are correct (refer 
to Section 2.1). Ensure the AC power 
switch is in the "|" (on) position and the 
RF ready light is illuminated. 

 

RF cable 
malfunction. 

Replace Model 1103 RF cable if using 
a Hand Held Sealing Head; contact an 
authorized Vante repair center if using 
a Bench Top Sealing Head. 

 

Problem with 
Sealing Head. 

See Sealing Head Instruction Manual. 

 

Blown Fuses. 

Ensure the fuse(s) is (are) not blown.  
See Section 4.2 if the fuse requires 
replacement. 

 

Component failure. 

Contact an authorized Vante

 repair 

center. 

Arcing or bad seals.  Contaminated 

sealing area. 

Clean sealing jaws (see Sealing Head 
Instruction Manual). 

 

Moisture in the 
sealing area or on 
the tubing. 

Ensure the outside of the tubing, the 
sealing region and adjacent areas are 
clean and dry. 

 

Premature release of 
hand held lever. 

Ensure the sealing indicator light is 
extinguished for at least two seconds 
prior to releasing the Hand Held sealing 
head lever. 

 

Misuse of sealing 
head. 

Ensure the sealing head is being used 
properly (see Sealing Head Instruction 
Manual). Ensure tubing is within the 
tubing specifications recommended 
within this Instruction Manual. Some 
other tubing materials may give poor 
seals or none at all. 

 

Sealing head jaws 
not aligned. 

Ensure the sealing head jaws close 
evenly (see Sealing Head Instruction 
Manual).  If not aligned, contact an 
authorized Vante

 repair center. 

 

Table 3.1  Troubleshooting Guide 

Summary of Contents for 4600

Page 1: ...Doc No 46000810 01 Rev D INSTRUCTION MANUAL MODEL 4600 RF Tube Sealer Power Source...

Page 2: ...ii This page intentionally left blank...

Page 3: ...9 12 Vante iii Vante 3480 E Britannia Dr Suite 120 Tucson AZ 85706 USA Telephone 520 881 6555 Fax 520 323 9055 Toll Free 877 565 5557 Part No 46000810 01 Rev D...

Page 4: ...iv This page intentionally left blank...

Page 5: ...odel 4605 4105 sealings heads it is the Power Source used to make seals on tubing made of RF reactive thermoplastic materials typically used in biopharmaceutical manufacturing Tubing utilized in the m...

Page 6: ...A Warning indicates a condition or procedure that could result in improper tube sealing or possible injury to the user A Warning is enclosed with a bold line box CAUTION A Caution indicates a conditi...

Page 7: ...tic Compatibility Directive 89 336 EEC Low Voltage Directive 73 23 EEC as amended by 93 66 EEC Conformity Assessment Body TUV SUD America For inquires related to the CE marking of this product please...

Page 8: ...viii This page intentionally left blank...

Page 9: ...s 5 1 6 Seal Cycle 6 1 7 Specifications 6 2 Power Source Operation 7 2 1 Setup 7 2 2 Sealing Procedure with Model 4105 Hand Held Sealing Head 8 2 3 Sealing Procedure with Model 4605 Bench Top Sealing...

Page 10: ...7 Figure 2 2 Model 4605 Bench Top Sealing Head Cable Attachment 8 Figure 2 3 Correct Technique for Holding the Sealing Head 8 Figure 2 4 Incorrect Technique for Holding the Sealing Head 9 Figure 2 5 M...

Page 11: ...xi Table of Tables Table 1 1 Tube Sealer Component Functions 3 Table 1 2 Tubing Specifications 5 Table 3 1 Troubleshooting Guide 16...

Page 12: ......

Page 13: ...it to dielectrically heat at a molecular level in the presence of RF energy This energy causes the plastic to soften due to the friction of the vibrating molecules In this softened condition the plast...

Page 14: ...pled with the various sealing heads becomes a system Figure 1 2 shows the component parts of the sealing system with a Hand Held Sealing Head and a Bench Top Sealing Head Table 1 1 indicates the funct...

Page 15: ...ble Connects the Model 4105 Sealing Head to the Power Source at the RF power output connector located on the front of the unit Model 4605 Cables Connects the Model 4605 Sealing Head to the Power Sourc...

Page 16: ...ower Source Power Interface Panel Figure 1 4 Serial Plate and Fuse Rating Labels for Power Source CAUTION Unit is equipped with dual fuses Disconnect power cord before replacing fuses RF Power Output...

Page 17: ...ance it is recommended that the Sealer be acclimated to the ambient environmental operating temperature for approximately one hour prior to use 1 5 Tubing Specifications Table 1 2 illustrates typical...

Page 18: ...wer on so the fan will continue to run Allow five minutes for cooling ensure the green seal ready light is on and resume sealing 1 7 Specifications Mechanical Dimensions Power Source 10 5 L x 4 4 W x...

Page 19: ...center Place the Power Source on a desk or lab counter for convenient use and route the AC power cord to prevent interference with other equipment or activities Verify the AC power switch is in the of...

Page 20: ...ocker type power switch on the rear of the Power Source to the on position The RF ready light on the front of the Power Source will illuminate immediately indicating that the Power Source is ready for...

Page 21: ...become damaged by arcing caused by contamination of the sealing region WARNING Do not place fingers in close proximity to the sealing region while sealing Radio frequency burns may occur Figure 2 4 I...

Page 22: ...guishes At that time the lever may be released Another seal cannot be made until the moveable lever has been fully opened See Figure 2 6 For more information refer to the Instruction Manual for the Mo...

Page 23: ...caused by contamination of the sealing region 2 3 Sealing Procedure with Model 4605 Bench Top Sealing Head NOTE For clarity the splash guard is not shown in some of the figures in this manual however...

Page 24: ...ject the operator to a radio frequency burn hazard Figure 2 8 Incorrect Sealing Method The seal is completed in approximately one second as indicated when the sealing jaw retracts Another seal cannot...

Page 25: ...ance may cause a rupture of a sealed tubing segment NOTE The Power Source is equipped with a thermal protection switch to prevent the equipment from becoming damaged If thermal cutout is reached the p...

Page 26: ...ess along a close ended length of tubing as in a segmenting process must be spaced no less than the bag manufacturer s recommendation to avoid rupture of the tubing due to pressure build up Multiple s...

Page 27: ...y kind of liquid This will damage the RF jaw tube assembly and void the warranty 3 1 Cleaning Power Source Case Ensure that the AC power switch located on the rear of the Power Source is in the off po...

Page 28: ...placement Component failure Contact an authorized Vante repair center Arcing or bad seals Contaminated sealing area Clean sealing jaws see Sealing Head Instruction Manual Moisture in the sealing area...

Page 29: ...Power Source may be the cause of malfunction In any event contact your nearest Vante authorized service representative or Vante for evaluation and repair 4 2 Fuse Replacement If the unit is plugged in...

Page 30: ...AC power cord into the input power receptacle and the AC outlet Turn the AC power switch to the on position If the RF Ready light fails to illuminate or if the fuse blows again contact Vante 4 3 Repai...

Page 31: ...nt shock protection Place them in a shipping carton large enough to contain the individually wrapped components or place each one in its own shipping carton of suitable size CAUTION Failure to properl...

Page 32: ...Vante Model 4600 20 This page intentionally left blank...

Page 33: ...n result in severe RF burns 5 3 RF Effects On Pacemakers There is no evidence that Vante RF instruments interfere with the function of modern cardiac pacemakers 5 4 Electrical Safety Vante RF devices...

Reviews: