background image

PURE BREEZE FUME HOOD

120V

25799 Jefferson Ave.

  

  

Murrieta, CA 92562

 

  

1-800-VANIMAN (826-4626)

  

  

www.vaniman.com

  

 

 

[email protected]

10370-11/21 PAGE 5

Page 5

MAINTENANCE

Before performing any maintenance tasks on this unit, flip the On/Off switch into 

the 

Off

 position and remove the supply cord from the unit.

1.  Filter Maintenance

 is most important to maintain a clean working environment. As dust, odors, and 

microbes are collected from the air in the room, the filters start to fill. When filters are full, they restrict 

air flow diminishing the efficiency of the 

Fume Hood

. Filters must be replaced regularly to assure the 

proper function of this unit.

a. 

The 

Pre-filter

 (first filter) is identified by its white media and is located on the very bottom of the 

stack. This filter captures the majority of larger sized contaminates in the room and will need to be 

changed more often than the other filters in the unit. Replace the Pre-filter when the air flow drops 

significantly, the filter appears substantially dirty, or every 3-6 months, which ever comes first. 

Replacing the Pre-filter: 

You will notice a group of three filters; the Pre-filter is the one on the bottom. In order to change 

the Pre-filter you must remove the two above it. Start removing filters from the top and work down 

through the group of filters. When installing the new Pre-filter make sure the metal reinforcement 

side is facing up. Inspect the other filters and replace them in the opposite order then they were 

removed.  

Order P/N 97021.

b. 

The 

Secondary Filter 

is identified by its white media and is the second filter from the bottom in 

the stack.This filter captures the smaller particles that may slip through the Pre-filter. As a rule, this 

filter should be replaced once every three replacements of the Pre-filter.     

Replacing the Second Filter:

Comparable to replacing the Pre-filter, you will need to remove the filter above first, then remove 

the Secondary filter. When installing the new Secondary filter, make sure the metal reinforcement 

side is facing up. 

Order P/N 97022

For your convenience Vaniman offers a filter package which includes three Pre-filters and one 

Secondary Filter. This filter package will enable you to replace your Pre-filters while reminding you 

of the three to one filter maintenance schedule. 

Order P/N 97025

.

c. 

The

 Charcoal Odor Filter 

(third filter) can be identified by its metal frame. The Charcoal Odor 

Filter is located on top of the group of filters inside the unit. This filter absorbs odors in the air as it 

is similiar to a chemical sponge used to collect odorous chemicals. The Charcoal Odor Filter must 

be changed when an increase of odors from the unit is detected.  

Repalcing the Charcoal Odor Filter: 

Pull out the old Charcoal Odor filter from the unit. Remove the plastic packaging from the new filter 

before sliding into place. Either side of the filter may face up. Vaniman recommends to always have 

a spare, however, keep new unused filters wrapped in their plastic wrappers to ensure freshness 

until use. 

Order P/N 97023.

Summary of Contents for VMC-A450

Page 1: ...Hood be sure that the location allows the power cord to be accessible at all times Additionally since the Pure Breeze Fume Hood draws air into the unit from the bottom and exhausts clean air from the...

Page 2: ...2 Laying the Fume Hood flat with the control panel facing up unlatch the door from the left side and open Remove all filters from the unit and remove the plastic bags covering each filter The plastic...

Page 3: ...each side for proper exhausting ventilation and access to the door latches i One piece of the bracket is mounted on the back of the unit at the factory ii Mount the second bracket piece with the open...

Page 4: ...he selector switch on the bottom of the unit to either low medium or high 6 If you are using Vaniman s Remote Switch accessory P N VMC A450 test it with the Fume Hood by plugging it into the socket la...

Page 5: ...filters and replace them in the opposite order then they were removed Order P N 97021 b The Secondary Filter is identified by its white media and is the second filter from the bottom in the stack This...

Page 6: ...the On Off switch into the Off position unplug the power cord from the unit correct the problem then push the breaker reset button until it clicks 2 Thermal Protection The motor in the unit is protect...

Page 7: ...vice disconnecting means cannot be locked securely fasten a prominent warning device such as a tag to service the panel c When servicing or replacing components unplug the power cord to ensure that th...

Reviews: