P3
– Faulty Combined PCB Air Switches
IMPORTANT
NOTE:
There are no serviceable parts on the Combined PCB. These air switches
are not serviceable and the whole board must be replaced if they fail.
As this is part of the
pneumatic system, it is important that all connections are airtight
.
Please note that all lifts will
come with new combined (Main + Aux.)PCB’s, C-450/C-625 Manual has only 2 air switches on
PCB, C-450/C-625 Power has 4 air switches on PCB and C-450/C-625 Power X-Y or Power
Turntable has 6 air switches on PCB.
1.
If there are no leaks in the hand control, airline tubing and grommets, there could be a leak
in air tubes to the pressure switch or failure of air switch. Check by performing the air leak
test with the hand control buttons.
2.
If there is an air leak,
tighten all air tubes and recheck.
Front side of new combined PCB (Main + Aux.) with air switches
Up
Down
Travers-
Gantry-
C-450/C-625 Technical Manual
031417
Page 15
Summary of Contents for C450
Page 11: ...C 450 C 625 Technical Manual 031417 Page 11 ...
Page 36: ...C 450 C 625 Technical Manual 031417 Page 36 ...
Page 38: ...C 450 C 625 Technical Manual 031417 Page 38 ...
Page 39: ...C 450 C 625 Technical Manual 031417 Page 39 ...
Page 40: ...l C 450 C 62 Technical C 450 C 625 Technical Manual 031417 Page 40 ...
Page 41: ...C 450 C 625 Technical Manual 031417 Page 41 ...
Page 42: ...C 450 C 625 Technical Manual 031417 Page 42 ...
Page 43: ...C 450 C 625 Technical Manual 031417 Page 43 ...
Page 44: ...C 450 C 625 Technical Manual 031417 Page 44 ...
Page 45: ...C 450 C 625 Technical Manual 031417 Page 45 ...
Page 46: ...17 C 450 C 625 Technical Manual 031417 Page 46 ...