Mechanical Systems
7.
The charger system does not work
The C-450/C-625 lift charger system has five components: a charger, charger end stop,
charger plates, wiring harness and circuit board.
Check that power is coming into the charger: the indicator light on the charger should
be green when the lift is not parked in the charging station.
Check that the charging strips of the charging station (bent in a wave pattern with the
middle section tensed and clear) make contact with the stainless steel contacts on the
trolley block.
Check the plugs along the wiring harness. The black wire from the charger should
correspond to the black wire from the harness.
Check the board by plugging the white plug into a replacement board.
8.
The on/off buttons (and/or emergency lowering buttons) on the hand control and the
lift do not work
Simultaneous failure of the pneumatic and electrical systems, strongly indicate a circuit
board failure. Confirm this by hooking a replacement board. In the case of emergency
lowering, if the replacement board does not resolve the problem, the fault is in the
“slack tape” lower limit switch. (See
point #1 ‘If the lift goes down to the end of the strap
and then goes up again with up and down functions reversed.’)
C-450/C-625 Technical Manual
031417
Page 10
Summary of Contents for C450
Page 11: ...C 450 C 625 Technical Manual 031417 Page 11 ...
Page 36: ...C 450 C 625 Technical Manual 031417 Page 36 ...
Page 38: ...C 450 C 625 Technical Manual 031417 Page 38 ...
Page 39: ...C 450 C 625 Technical Manual 031417 Page 39 ...
Page 40: ...l C 450 C 62 Technical C 450 C 625 Technical Manual 031417 Page 40 ...
Page 41: ...C 450 C 625 Technical Manual 031417 Page 41 ...
Page 42: ...C 450 C 625 Technical Manual 031417 Page 42 ...
Page 43: ...C 450 C 625 Technical Manual 031417 Page 43 ...
Page 44: ...C 450 C 625 Technical Manual 031417 Page 44 ...
Page 45: ...C 450 C 625 Technical Manual 031417 Page 45 ...
Page 46: ...17 C 450 C 625 Technical Manual 031417 Page 46 ...