5.
Insert the cables to their grooves in the 4-position lead-through packing and make
sure the rubber seals are correctly in position. Push the packing into its hole in the
correct position, so that its groove matches with the guide pin (see arrows in
figure below), and tighten it enough to make sure that the cables will not slip
through and the connection is tight.
6.
Position the electronics and press gently until the mounting screw heads are
aligned with the threaded holes on the base plate. First turn the mounting screws
(4 pcs) by hand, and then tighten them with a screwdriver when you are sure that
they go correctly into the threads.
7.
Connect the power, pressure sensor and temperature sensor cables to
electronics.
8.
Connect the antenna cables to electronics. Tighten the connectors with the SMA
torque wrench to torque 1 Nm.
Do not over-tighten!
9.
Connect the sensor cable. Make sure that the green LED begins to blink,
indicating that the software is running.
10.
Install the sensor cover.
11.
Calibrate the sensor.
Replacing parts | 77 (93)
K17122 V1.1 EN - 1 - 06/2020
Summary of Contents for K17122
Page 34: ...10 3 Menu map Operating 34 93 K17122 V1 1 EN 1 06 2020...
Page 58: ...Advanced calibration 58 93 K17122 V1 1 EN 1 06 2020...
Page 60: ...Advanced calibration 60 93 K17122 V1 1 EN 1 06 2020...
Page 88: ...18 WEEE information Valmet Bridge Sensor WEEE information 88 93 K17122 V1 1 EN 1 06 2020...
Page 91: ...2 3 Valmet MCA Spare parts Valmet Automation Inc All rights reserved 7 10 9 12 14 15 13 11...
Page 92: ...3 3 Valmet MCA Spare parts Valmet Automation Inc All rights reserved 16 17...