8.
Disconnect the antenna cables with the SMA torque wrench.
Installing:
1.
Install new antenna cables and tighten with the SMA torque wrench.
2.
Push cables through the base plate and press sensor body back on the
measurement head.
3.
Turn the sensor upright, put base plate on sensor body and align the long
mounting bolts to their holes in the measurement head. Make sure that the
sealing washers are in position. Apply some Loctite 243 locking glue to the screw
threads.
4.
Tighten the bolts, torque about 19 Nm.
Replacing parts | 76 (93)
K17122 V1.1 EN - 1 - 06/2020