8 Operating terminal
8.1 Mechanical installation
Note:
Do not mount the operating terminal to a structure that is subjected to
vibrations: for example a process pipe, pipe support, or any structure
connected to a process pipe.
When selecting a place for the operating ter-
minal, remember that terminal is delivered
with a 10 m (33 ft) sensor cable. If a longer
cable is needed, use one or more extension
cables (option; length 10 m / 33 ft). Maximum
cable length is 30 m / 99 ft. If the required ca-
ble is longer than 30 m, please contact your
Valmet representative for further advice.
Attach the operating terminal to the wall or
other non-vibrating location with three mount-
ing screws in a place that is easily accessed.
Suitable mounting level is 160–165 cm (5 ft
3–5") from the floor to the upper mounting
hole.
Operating terminal | 23 (93)
K17122 V1.1 EN - 1 - 06/2020