User’s Guide ______________________________________________________________________
230 _________________________________________________________________ M211840EN-C
Auxiliary
sensor
averaging
time
3
0.25 ... 3600 seconds
IU,A
Auxiliary
rain sensor
gain
0.2
0.000000001 ... 1000000
IA,G
Reset mode
(cumulative
rain amount
resetting)
M
M = resets rain when pulse
amount exceeds 65535. When
gain is 0.2, it resets at
65536*0.2 = 13107 (mm)
L = resets when reaching
limit value
A = Cumulative rain value is
reset every time rain
information is sent out from
transmitter (user counts
cumulative rain amount).
IA,M
Rain reset
limit
1000
0.000000001 ... 1000000
IA,L
Auxiliary
solar sensor
gain
100000
0.000000001 ... 1000000
IB,G
Auxiliary
level sensor
gain
1
0.000000001 ... 1000000
IS,G
Auxiliary
level sensor
operating
range
1
0 = 0 to 2.5 V range
1 = 0 to 5.0 V range
2 = 0 to 10.0 V range
IS,M
Auxiliary
temperature
averaging
1
0.5 ... 3600
IP,A
Table 54
Auxiliary Sensor Parameters
Parameter
Factory
Range
Info
Command