Chapter 6 ______________________________ Serial Line Communication With PTB330 Barometer
VAISALA_______________________________________________________________________ 75
SERI [b p d s]
Example:
>seri
Baud P D S : 9600 N 8 1
>
You need to reset the barometer to activate the new communication
settings set with the
SERI
command.
The settings can be changed one parameter at a time or all parameters
at once:
>seri 0
Baud P D S : 9600 N 8 1
>seri o
Baud P D S : 9600 O 8 1
>
SMODE
Use the command
SMODE
to set the user port start-up operating mode.
SMODE [xxxx]
where
b
=
Bit rate (110, 150, 300, 600, 1200, 2400, 4800, 9600,19200,
38400, 57600, 115200)
p
=
Parity (n = none, e = even, o = odd)
d
=
Data bits (7 or 8)
s
=
Stop bits (1 or 2)
where
xxx =
STOP, RUN, POLL, or TS