User's Guide ______________________________________________________________________
70 ___________________________________________________________________M210983EN-A
Example:
>avrg
Average filter : 1 s ?
>
DPMAX
Use the
DPMAX
command to show or set the maximum pressure
difference allowed between barometer modules. The command is only
available if more than one barometer module has been installed.
DPMAX [x]
Example:
>dpmax 0.2
Max. diff. : 0.20 hPa
>
TQFE
Use the
TQFE
command to show or set the temperature for QFE
corrected pressure. The valid range of QFE temperature is -80...+200
°C.
Example:
>tqfe
QFE temp. : 20.00 'C ?
>
where
x
=
Pressure