Property
Description
Initialization time to full accuracy
Two days
Data storage
At least 10 years
Expected operating life
> 15 years
1)
Three cycles for ethane and hydrogen.
Table 13 Calculated Parameters
Property
Description
Total dissolved combustible gases
(TDCG)
Combined total of H
2
, CO, CH
4
, C
2
H
6
, C
2
H
4
, and C
2
H
2
Rate of change (ROC)
Available for single gases and TDCG for 24 h, 7 d, and 30 d
periods
Gas ratios
Available ratios:
• CH
4
/H
2
• C
2
H
2
/C
2
H
4
• C
2
H
2
/CH
4
• C
2
H
6
/C
2
H
2
• C
2
H
4
/C
2
H
6
• CO
2
/CO
1)
Calculated from 24 h average values. See standard IEC 60599.
Table 14 Operating Environment
Property
Description
Transformer oil type
Mineral oil
Required minimum fire point
of
transformer oil
+125 °C (+257 °F)
Transformer oil pressure at oil inlet
Max. 2 bar
abs
continuous
Burst pressure 20 bar
abs
Transformer oil temperature at oil inlet Max. +100 °C (+212 °F)
Ambient humidity range
0 ... 100 %RH, condensing
Ambient temperature range in
operation
-40 ... +55 °C (-40 ... +131 °F)
OPT100 User Guide
M211858EN-E
54
Summary of Contents for Optimus OPT100
Page 1: ...M211858EN E User Guide Vaisala Optimus DGA Monitor for Transformers OPT100...
Page 24: ...Figure 11 Measurement Graph for 1 Day Rate of Change Values OPT100 User Guide M211858EN E 22...
Page 72: ...OPT100 User Guide M211858EN E 70...
Page 76: ...OPT100 User Guide M211858EN E 74...
Page 77: ......
Page 78: ...www vaisala com...