Register Number
(Decimal)
Address
(Hexadecimal)
Register
Description
Data Format
Unit and Notes
Gas Ratios (24h Average Values)
141
008C
hex
CH
4
/H
2
32-bit float
-
143
008E
hex
C
2
H
2
/C
2
H
4
32-bit float
-
145
0090
hex
C
2
H
2
/CH
4
32-bit float
-
147
0092
hex
C
2
H
6
/C
2
H
2
32-bit float
-
149
0094
hex
C
2
H
4
/C
2
H
6
32-bit float
-
151
0096
hex
CO
2
/CO
32-bit float
-
Other
161
00A0
hex
Time stamp of
measurement
result set, in
seconds since
start of Unix
epoch
64-bit float
(double)
s
165
00A4
hex
Index of
measurement
result set
32-bit integer
-
A.3.3 Alert Registers
Gas level alert registers are 16-bit fields. Each register communicates the status of alerts of
specific severity for a particular timeframe (for example, gas level caution alerts for present
values).
If an alert is active, the corresponding bit is set to 1. The bit is set to 0 when the alert is
acknowledged or the alert is turned off. For more information, see
.
OPT100 User Guide
M211858EN-E
64