9.
Use the arrow buttons to select
ECHO
. Press
ON
to turn to it on.
Press
OFF
to turn it off.
10. Press
EXIT
to return to the basic display.
SERI
Use the serial line command
SERI
[
b p d s
] to set communication
settings for the user port.
Syntax:
SERI
[
b p d s
]
where
b
=
Bauds (110, 150, 300, 600, 1200, 2400, 4800, 9600,19200,
38400, 57600, 115200)
p
=
Parity (n = none, e = even, o = odd)
d
=
Data bits (7 or 8)
s
=
Stop bits (1 or 2)
Example:
>SERI 600 N 8 1
600 N 8 1
>
SMODE
Use the command
SMODE
[
xxxx
] to set the user port start-up operating
mode.
Syntax:
SMODE
[
xxxx
]
where
xxxx
=
STOP, RUN, POLL or SEND