Chapter 4 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 75
3.
Open a terminal program and set the communication settings as
follows:
Table 14
Communication Settings for the Service Port
Parameter Value
Bauds 19200
Parity None
Data bits
8
Stop bits
1
Flow control
None
For a detailed explanation of using a terminal program, see section
Terminal Program Settings on page 82.
4.
Power-up the DMT346.
LAN Communication
To enable LAN communication, a LAN or WLAN interface must be
physically connected to the network, and the networking settings must be
suitable for your network. For a description of interfaces, see sections
LAN Interface on page 56 and WLAN Interface on page 57.
The LAN and WLAN interfaces both operate by accessing the serial
interface (User Port) of the transmitter. All commands that are available
using the serial interface are available through the LAN and WLAN
interfaces; refer to section List of Serial Commands on page 85. For
instructions on how to connect using a terminal program, see section
Terminal Program Settings on page 82.
IP Configuration
The IP settings of the LAN and WLAN interfaces are described in Table
15. The current settings can be viewed on the serial line or using the
device information display; see section Device Information on page 98.