USER'S GUIDE____________________________________________________________________
134 _________________________________________________________________ M210762EN-D
5.
Wait at least 30 minutes for the sensor to stabilize. Follow the
stabilization from the
GRAPH
display. Press
EXIT
to return to the
previous display.
6. Press
READY
when stabilized. Enter the reference value by using
the arrow keys.
0706-019B
Figure 68
Waiting for Readings to Stabilize
Now proceed to the adjustment at the wet end reference condition
(30 ... 75 %RH) and carry out the procedure as described in the
previous items.
7. Answer
YES
to confirm the adjustment. Press
OK
to return to the
adjustment menu.
8.
Before closing the adjustment mode, feed the adjustment
information into the device, see section Adjustment Information on
page 132. Press
EXIT
to close the adjustment mode and return to
the basic display.
Using Serial Line
NOTE
Reference humidities must be ~0 %RH and 30 ... 75 %RH. A suitable
product for the two point RH calibration is for example the Vaisala
Humidity Calibrator HMK15.
1.
Connect the DMT345/346 to a PC. See section Serial Line
Communication on page 72. Open a terminal program.
2.
Carry out the chemical purge.
3. Press
the
ADJ
button.
4.
Remove the filter from the probe and insert the probe in the dry end
reference conditions (~0 %RH).
5. Issue
the
FCRH
command and press
Enter
.
FCRH
6.
Wait at least 30 minutes for the sensor to stabilize.
7. Press
Enter
a few times to check if the reading is stabilized.