![Vaisala CAB100 Installation Manual Download Page 115](http://html2.mh-extra.com/html/vaisala/cab100/cab100_installation-manual_3171839115.webp)
Property
Description/Value
Overpressure
Proof pressure
Burst pressure
Static pressure
1.0 bar
1.7 bar
1.7 bar
Mounting position
Error (zero adjustable)
≤
1 %/g (calibration in vertical position is standard)
Adjustments (front accessible)
Zero
Span
±5 % span
±3 % span
Table 11 PDT101 operating environment
Property
Description/Value
Operating temperature
−18 ... +70 °C (−0.4 ... +158 °F)
Storage temperature
−40 ... +82 °C (−40 ... +179.6 °F)
EMC compliance
EN 61326-1, Basic immunity test requirements
Note: If used in an electromagnetic field of 3 V/m, with narrow frequency area of 80 ... 120 Mhz, it is possible that the current output of PDT101 can
deviate max. 0.8 % (with accuracy specified 0.4 %)
Table 12 PDT101 inputs and outputs
Property
Description/Value
Process connection
1/4" barbed fittings
Max. loop resistance for 4 ... 20 mA
≤
(Supply voltage - 12 V)/0.022 A
Supply current
Max. 20 mA for 4 ... 20 mA output signal
Optical process diagnostics
LED visual indicator
Electrical connection
Euro style pluggable terminal block accepts 12 ... 26
AWG wire
(0.13 up to 3.31 mm
2
)
Output signal
2-wire
3-wire
4 ... 20 mA
0 ... 5 VDC (user selectable 0 ... 10 VDC)
Operating voltage
2-wire output 4 ... 20 mA
3-wire output 0 ... 5 VDC
3-wire output 0 ... 10 VDC
12 ... 36 VDC
11.5 ... 36 VDC
14 ... 36 VDC or 24 VAC
Chapter 9 – Technical data
113
Summary of Contents for CAB100
Page 1: ...M212242EN B Installation Guide CMS Industrial Cabinet CAB100...
Page 144: ......