During operation
➨
Vent and dispose of
potentially dangerous gases or
vapors
at the outlet of the pump appropriately.
+
Due to the high compression ratio, the pump might gen-
erate overpressure at the outlet. Check pressure com-
patibility with system components (e.g., exhaust tubing
or exhaust valve) at the outlet. Ensure that the pump
outlet is neither blocked nor restricted.
+
Maximum ambient temperature
: 104 °F (40 °C)
Check the maximum temperatures, if installing the
pump in a cabinet or a housing. Make sure ventilation
is adequate, especially if the ambient temperature is
elevated.
+
Operation with silencer at the outlet: Operating the
pump at a high inlet pressure or pumping dusty gas-
es for a long time may cause clogging of the silencer.
Check the silencer regularly and replace if necessary.
• If the pump is installed at an altitude of more than 3300
ft (1000 m) above mean sea level, check compatibil-
ity with applicable safety requirements, especially IEC
60034. There is a risk of the motor overheating due to
insufficient cooling.
• Check compatibility with the
maximally permitted
pressure
at outlet and the
maximum pressure differ-
ence
between inlet and outlet ports.
If pumping condensable vapors (water vapor, solvents,
etc.), let the pump run with
gas ballast
to help purge any
condensation in the pump.
Do not start the pump if the
pressure difference between
inlet and outlet ports exceeds max. 16.0 psi (1.1 bar)
.
Attempts to start the pump at higher pressure difference
may cause stalling and damage of the motor.
Prevent internal condensation, transfer of liquids or dust.
NOTICE