page 14 of 57
the setting of the voltage selection switch while the
pump is connected to AC power. Unplug the pump be-
fore setting the voltage selection switch.
Note
: If the
pump is switched on with wrong voltage selection, the
motor may be damaged!
Keep a minimum distance of 8 in (20 cm) between the
cooling fan and surrounding items (e.g., housing, walls,
etc.). Check fan regularly for dust/dirt. Clean fan guard
grill if necessary to avoid a reduction of ventilation.
Use only hoses at the inlet and outlet of the pump with
an inner diameter at least as large as the diameter of the
pump’s fittings (to avoid overpressure at the outlet, and
reduction of pumping speed at the inlet).
Allow the equipment to equilibrate to ambient temperature
if you bring it from cold environment into a room prior to
operation. Notice if there is water condensation on cold
surfaces.
Comply with all
applicable and relevant safety require-
ments
(regulations and guidelines).
Implement the re-
quired actions and adopt suitable safety measures.
Ambient conditions
➨
Do not reach for this product if it has fallen into liquid.
There is a risk of deadly electrical shock. Unplug the
system immediately.
+
Do not use this product in an area where it can fall or
be pulled into water or other liquids.
+
Pay attention to the
permissible maximum ambient
and gas inlet temperatures
(see ”Technical data”, pg.
NOTICE